510(k)K220161

FDA DEVICE FACTS

Biofeedback Nerve and Muscle Stimulator

Shenzhen Konmed Technology Co., Ltd.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJuly 20, 2022PRODUCT CODEKPI

Indications for use

The device is intended to provide electrical stimulation and neuromuscular re -education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women. As a powered muscle stimulator the Biofeedback Nerve and Muscle Stimulator is indicated for the following conditions: ·Relaxation of muscle spasm ·Prevention or retardation of disuse atrophy ·Increasing local blood circulation ·Muscle re-education ·Immediate post -surgical stimulation of calf muscles to prevent venous thrombosis ·Maintaining or increasing range of motion As a biofeedback device the Biofeedback Nerve and Muscle Stimulator is indicated for the following conditions: ·Biofeedback, relaxation and muscle re-education purposes As a nonimplanted electrical continence device the Biofeedback Nerve and Muscle Stimulator is indicated for the following conditions: ·Acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Inhibition of the detruser muscles through reflexive mechanisms and strengthening of pelvic floor muscles. ·Incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles such as the abdominal and the gluteus muscles. Prescription or OTC OTC RX Only

Power source 7.4V DC/1200mAh rechargeable lithium battery 7.4V DC/1200mAh rechargeable lithium battery Maximum Output Current (±10%) 94.4mA @ 500Ω 54mA @ 2kΩ 15 mA @ 10kΩ 94.4mA @ 500Ω 54mA @ 2kΩ 15mA@ 10kΩ Maximum Phase Charge 51.4µC@500Ω 51.4µC @ 500Ω Maximum Current Density 6.01mA/ cm2@ 500Ω Surface = 15.7 cm² 6.01mA/ cm2 @500Ω Maximum Power Density 0.2814mW/ cm2@ 500Ω 0.2814mW(12mW) / cm2@ 500Ω

Frequency (Hz) 2-100Hz 2 -100Hz

Pulse Duration (µsec) 50-450μs 50- 450 μs Net charge For pulsed symmetric, biphasic: 0μC @ 500Ω; For pulsed symmetric, biphasic: 0μC @ 500Ω; As evidenced by the above table, both the subject and the predicate devices have the same intended use. Additionally, the subject device has identical technological characteristics to the predicate except the predicate device was cleared for prescription only use and the subject device is an over the counter (OTC) use device. However, th is difference do es not raise different questions of safety or effectiveness and the testing mentioned below showed that the subject is substantially equivalent with the predicate.

See the full product information製品情報をすべて見るには