FDA DEVICE FACTS
HD+12, HD+15
Cardioline S.P.A
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The function of the device is the acquisition and transmission of the ECG signal in order to support diagnosis of the patient’s conditions. HD+ is a wireless or USB (with appropriate option) ECG acquisition device, to be used primarily as common front-end for standard PC/tablet platforms (Windows/MAC OS/others), for resting ECG applications. The device implements wireless communication via Bluetooth wireless technology or wired with USB communication. With both connection modes, HD+ sends the data to the receiver device without performing any analysis or filtering.
HD+ is not intended for monitoring or analysis of the cardiac function or to diagnose the patient’s health condition. Display, print and analysis applications on the receiver device are separate products. HD+ is not able to permanently store the acquired data, therefore it does not work unless a connection has been established with a receiver application. Furthermore, HD+ does not collect any of the patient’s sensitive data (patient’s name, age, previous health conditions etc.).
HD+ detects the QRS complexes and transmit the results to the receiving device. The QRS detection function is intended for patients aged 12 years or older.
▪ HD+ is indicated for the acquisition of ECG signals with the patient in resting conditions, for example for diagnostic ECG’s and rhythm strips
▪ HD+ is suitable for working at high altitudes, with restrictions.
▪ HD+ is intended for use on patients, with no limits of age or gender, except for the HR detector function which is intended for patients 12 years or older
▪ HD+ is intended for use in medical facilities (hospitals, clinics), at home or emergency settings (ambulances).
▪ HD+ is intended for use by a physician, nurse or other healthcare professional (e.g., ECG technician) who acts following orders by a physician or authorized nurse, including when operated in home environments
▪ HD+ is not intended for real-time monitoring of vital physiological parameters.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。