Indications for use
The LimFlow ARC is intended to facilitate placement and positioning of guide wires and catheters within the peripheral vasculature. The LimFlow ARC is not intended for use in the coronary or cerebral vasculature.
Comparison of Technological Characteristics with the Predicate Device The LimFlow ARC is substantially equivalent to the predicate device based upon the following similarities:
• Both devices are intended to be used to cross chronic total occlusions (CTOs) in peripheral vessels
• Both devices are used in cardiac catheter labs in either a hospital or an office-based lab
• Both devices are advanced to the target occlusion through an indwelling vascular sheath
• Advancement of both devices is monitored by external fluoroscopy
• Both devices have equivalent sizes in terms of outer diameter and working length of the cannula
• Both devices share the same basic design, mechanism of action, and principles of operation
• Both devices share the same sterilization method (EO) The table below compares the LimFlow ARC (this submission) to the predicate device, the Outback Elite (K150836).
Device Feature OUTBACK Elite (Predicate) LimFlow ARC (This Submission) Device Classification Class II Same Classification Name Catheter for Crossing Total Occlusions (21 CFR 870.1250) Same Device Category Percutaneous catheter Same Indications for Use The OUTBACK Elite is intended to facilitate placement and positioning of guide wires and catheters within the peripheral vasculature. The OUTBACK Elite is not intended for use in the coronary or cerebral vasculature. Same Device Description: Design & Operating Principle The OUTBACK Elite is a single-use device designed to facilitate placement and positioning of guide wires within the peripheral vasculature. The device consists of three primary elements: 1) Cannula, 2) Catheter shaft, and 3) Deployment handle with deployment control slide. The OUTBACK Elite Re- Entry Catheter is supplied sterile and is available in two usable lengths (80 cm and 120 cm). Same, except the is available in one usable length (100 cm). Packaging Contents One (1) OUTBACK Elite Same Device Feature OUTBACK Elite (Predicate) LimFlow ARC (This Submission) Contraindications The OUTBACK Elite is not intended for use in the coronary or cerebral vasculature. Same Where used Hospital or office-based lab Same