AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, V-PRO 60 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector

Steris

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATESeptember 30, 2022PRODUCT CODENVE

Indications for use

The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles (Table 1): Table 1. AMSCO 600 Steam Sterilizer factory-validated sterilization cycles and cycle values Cycles Sterilize Temperature Sterilize Time Dry Time Maximum Recommended Load Prevac 270°F (132°C) 4 minutes 20 minutes Fabric Packs. Refer to Table 2 for recommended quantities. Prevac 270°F (132°C) 4 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each and Fabric Packs. Refer to Table 2 for recommended quantities. Prevac 270°F (132°C) 10 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each . Refer to Table 2 for recommended quantities. Prevac 270°F (132°C) 4 minutes 5 minutes Single Fabric Pack. Prevac 275°F (135°C) 3 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. Prevac- IUSS 270°F (132°C) 4 minutes 1 minutes Immediate use – single unwrapped tray Gravity 250°F (121°C) 30 minutes 30 minutes Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. Warm-Up 270°F (132°C) 3 minutes 1 minute N/A DART 270°F (132°C) 3.5 minutes 1 minute Bowie-Dick Test Pack, DART Leak Test N/A N/A N/A N/A Table 2 AMSCO 600 Steam Sterilizer full load per sterilizer size Sterilizer Size Wrapped Instrument Trays Fabric Packs 26” x 26” x 39” 9 12 26” x 26” x 51” 12 16 26” x 26” x 63” 15 20 The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 600 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader. The V-PRO maX 2 Low Temperature Sterilization System using V APROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues. Each Cycle can sterilize non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and Fast Non Lumen Cycle. The Non Lumen Cycle can sterilize: lNon-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. l The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22. 7 kg). The Fast Non Lumen Cycle can sterilize: *Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. * The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg). The Flexible Cycle can sterilize: Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:

1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.* The flexible endoscopes may contain either:

• A single lumen that is� 1 mm internal diameter (ID) and :S 1050 mm in length

• Or two lumens with:

- One lumen that is � 1 mm ID and :S 990 mm in length

- And the other lumen that is � 1 mm ID and :S 850 mm in length * The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).

2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. tt The flexible endoscope may contain either: A single lumen that is � 1 mm ID and :S 1050 mm in length

• Or two lumens with:

- One lumen that is � 1 mm ID and :S 990 mm in length

- And the other lumen is � 1 mm ID and :S 850 mm in length. tt The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg). The Lumen Cycle can sterilize: tMedical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single channeled devices with a stainless lumen that is::=-: 0.77 mm ID and :S 500 mm in length

• Single channeled devices with a stainless lumen that is ::=-: 1.8 mm ID and :S 542 mm in length

• Dual channeled devices with stainless lumens that are::=-: 0.77 mm ID and :S 527 mm in length

• Triple channeled devices with stainless lumens that are either: ::=-: 1.2 mm ID and :S 275 mm in length ::=-: 1.8 mm ID and :S 310 mm in length or ::=-: 2.8 mm ID and :S 317 mm in length t Validation testing for all lumen sizes was conducted using a maximum of20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg). The V-PRO s2 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The pre-programmed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues. Each Cycle can sterilize non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. The V-PRO s2 Sterilizer Non Lumen Cycle can sterilize:‡ Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. ‡ The validation studies were conducted using a validation load consisting of one instrument tray for a total weight of 25 lbs (11.3 kg). The V-PRO s2 Sterilizer Fast Cycle can sterilize:‡ Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) internal diameter (ID) and ≤ 410 mm (~16-9/64”) in length ≥ 1.8 mm (~5/64") ID x ≤ 542 mm (~21-5/16") in length

• Triple channeled devices with stainless steel lumens that are either: ≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length ≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length or ≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. ‡ Validation testing for all lumen sizes was conducted using a maximum of eight (8) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. Validation testing was conducted using a validation load consisting of one pouched instrument tray and two pouched devices outside of the tray with a total weight of 4.0 lbs (~1.8kg). The V-PRO s2 Sterilizer Flexible Cycle can sterilize:@ One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

• The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm ID and ≤ 990 mm in length

• Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions ≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9 11/64”) in length ≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (~10”) in length ≥ 1.8 mm (~5/64”) ID and ≤ 542 mm (~21-5/16”) in length @ The validation studies were conducted using a validation load consisting of two instrument trays. One tray contained one flexible endoscope, with silicone mat, instrument organizers and light cord (if not integral to scope), and the second tray contained additional load and twelve (12) stainless steel lumens for a total load weight of 11 lbs (5 kg). Hospital loads should not exceed the maximum number of lumens validated by this testing. The V-PRO s2 Sterilizer Lumen Cycle can sterilize: ^ Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) ID and ≤ 410 mm (16-9/64”) in length ≥ 1.8 mm (~5/64) ID x ≤ 542 mm (~21-5/16") in length

• Triple channeled devices with stainless steel lumens that are either: ≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length or ≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length ^ Validation testing for all lumen sizes was conducted using a maximum of twelve (12) stainless steel lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray and two pouches for a total weight of 11 lbs (5.0 kg). The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si End oWrist® instruments. The V-PRO® 60 Low Temperature Sterilization System using VAPROX® HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The pre-programmed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues. Each Cycle can sterilize non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. The V-PRO 60 Sterilizer Non Lumen Cycle can sterilize:‡ Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. ‡ The validation studies were conducted using a validation load consisting of one instrument tray for a total weight of 25 lbs (11.3 kg). The V-PRO 60 Sterilizer Flexible Cycle can sterilize:@ One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

• The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm (~3/64”) internal diameter (ID) and ≤ 990 mm (38-63/64”) in length

• Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions ≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9-11/64”) in length ≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (~10”) in length ≥ 1.8mm (~5/64”) ID and ≤ 542 mm (~21-5/16”) in length @ The validation studies were conducted using a validation load consisting of two instrument trays. One tray contained one flexible endoscope, with silicone mat, instrument organizers and light cord (if not integral to scope), and the second tray contained additional load and twelve (12) stainless steel lumens for a total load weight of 11 lbs (5 kg). Hospital loads should not exceed the maximum number of lumens validated by this testing. The V-PRO 60 Sterilizer Lumen Cycle can sterilize: ^ Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are ≥ 0.77 mm (~1/32”) ID and ≤ 410 mm (16-9/64”) in length ≥ 1.8 mm (~5/64") ID x ≤ 542 mm (~21-5/16") length

• Triple channeled devices with stainless steel lumens that are either: 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length ≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length or ≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length ^ Validation testing for all lumen sizes was conducted using a maximum of twelve (12) stainless steel lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray and two pouches for a total weight of 11 lbs (5.0 kg).

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 22, 2026