510(k)K230095

FDA DEVICE FACTS

ScanX Swift 2.0, ScanX Swift View 2.0

Durr Dental SE

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEFebruary 6, 2023PRODUCT CODEMUH

Indications for use

The ScanX Edge is intended to be used for scanning and processing digital images exposed on Phosphor Storage Plates (PSPs) in dental applications. The ScanX Swift 2.0 is intended to be used for scanning and processing digital images exposed on Phosphor Storage Plates (PSPs) in dental applications. The ScanX Swift View 2.0 is intended to be used for scanning and processing digital images exposed on Phosphor Storage Plates (PSPs) in dental applications. Identical. Technical equivalence: Operating principle The PSPs are put into the mouth of the patient, exposed to X-rays, and then are read out with the device. The read-out-process is carried out with a 639 nm Laser. The laser beam stimulates the top coating of the plates, which consists of x-ray sensitive material. Depending on the exposed dose, the coating emits different levels of light. These light particles are then requisitioned by an optical sensor and transferred into an electrical output signal. This signal is digitalized and is the data for the digital X-ray image. The PSPs are put into the mouth of the patient, exposed to X-rays, and then are read out with the device. The read-out-process is carried out with a 639 nm Laser. The laser beam stimulates the top coating of the plates, which consists of x-ray sensitive material. Depending on the exposed dose, the coating emits different levels of light. These light particles are then requisitioned by an optical sensor and transferred into an electrical output signal. This signal is digitalized and is the data for the digital X-ray image. Identical. Page | 4

Device Predicate device Subject devices Change Impact Analysis Data transfer The data is transmitted via an Ethernet link. The data is transmitted via an Ethernet link to a computer. With the XPS07.2A1 the data can alternatively be transferred by a WLAN interface or being stored on a removeable storage medium. Similar. Erasing the residual image following scanning for plate reuse After the scanning process, the remaining data is erased by a LED light. After the scanning process, the remaining data is erased by a LED light. Identical.

Display The ScanX Edge does not have a screen / display.

The device state is indicated via status LEDs lights on the device.

2.4’’-Display

The display does not show preview images. 5’’-Touch Display: The touch screen only shows a preview which serves to provide an initial impression of the final x-ray image. For the purposes of diagnosis, the x-ray image must be viewed on a diagnostic PC monitor. Diagnostic monitor = PC monitor. The scanner itself is not used for diagnosis. Similar.

Imaging scanning Laser / Photomultiplier Tube (PMT) Components: Photomultiplier 2" Diode, Laser 639nm/10mW Fiber coupled laser diode Laser / Photomultiplier Tube (PMT) Components: Photomultiplier 2" Diode, Laser 639nm/10mW Fiber coupled laser diode Identical. Plate guide A specific plate guide for each PSP size is available, thus the user chooses which size of PSP he wants to use and prepares the device by inserting the appropriate plate guide into the device. A specific plate guide for each PSP size is available, thus the user chooses which size of PSP he wants to use and prepares the device by inserting the appropriate plate guide into the device. Identical. Page | 5

Device Predicate device Subject devices Change Impact Analysis Transport / feed mechanism The plates are inserted into a size specific interchangeable insert. If the user wants to scan a different plate size, the insert needs to be exchanged. The insert is used to transport the plate into the scanning position. During the scanning process the plate is not move. Upon completion of the scanning. The plates are inserted into a size specific interchangeable plate guide. If the user wants to scan a different plate size, the plate guide needs to be exchanged. The plate is transported beltway down the axis of the cylinder past the slot. The motion of the laser and plates provides the two orthogonal scan directions. This is a continuous feed device that allows successive plates to be loaded as soon as the previous plates have moved past the slot. Similar.

The solutions for the plate guides have evolved for ScanX Swift 2.0 and ScanX Swift View 2.0. The feed and transport are equivalent.

WLAN (Wi-Fi) No No Yes Different. Added interface flexibility Ethernet Yes Yes Yes SAME RFID Supports image plates with RFID Tags. The information on the RFID tag is used to identify the image plate. Supports image plates with RFID Tags. The information on the RFID tag is used to identify the image plate. Supports image plates with RFID Tags. The information on the RFID tag is used to identify the image plate. Identical.

The RFID image plates are identical. Page | 6

Device Predicate device Subject devices Change Impact Analysis Operation / Function The image plate scanner is used to read image data stored on an image plate and to transfer the data to the imaging software (e.g. VisionX) on a computer.

The transport mechanism guides the image plate through the device. The image plate is read using a laser inside the scanner unit. The scanned data is converted into a digital image and transferred to the imaging software.

After scanning, the image plate runs through the erasure unit. Image data still held on the image plate is erased with the aid of bright light. The image plate is then ejected for re-use. The phosphor storage plate scanner is used to scan image data stored on a phosphor storage plate and transfer the data to the imaging software (e.g. VisionX) on a computer.

The transport mechanism guides the phosphor storage plate through the device. The phosphor storage plate is read using a laser inside the scanner unit. The scanned data is converted into a digital image and transferred to the imaging software.

After scanning, the phosphor storage plate runs through the erasure unit. Image data still held on the phosphor storage plate is erased with the aid of bright light. The phosphor storage plate is then ejected for re-use. The phosphor storage plate scanner is used to read image data stored on the phosphor storage plate. The unit can be used in two different ways: via the imaging software (e. g. VisionX) on a PC or directly via the touch screen on the unit.

The transport mechanism guides the phosphor storage plate through the device. A laser in the scanner unit scans the PSP. The scanned data is converted into a digital image.

If a scanning job is started via the imaging software, the image is automatically transmitted to the computer. If a scanning job is started via the touch screen, the image is saved to the memory card and needs to be transferred to the computer later on. After scanning, the phosphor storage plate runs through the erasure unit. Image data still held on the phosphor storage plate is erased with the aid of bright light. The phosphor storage plate is then ejected for re-use. Identical.

Different: ScanX Swift View 2.0: The touch screen allows the unit to be operated in a stand-alone mode (without a connection to a computer). The device is able to generate the image data itself and to store it on a data medium (USB stick). Instructions can be entered on the touch screen with the tip of a finger.

Phosphor storage plates Only AT Phosphor Storage Plates with RFID tag: Size 0: 22 x 35 mm Size 1: 24 x 40 mm Size 2: 31 x 41 mm

Only AT Phosphor Storage Plates with RFID tag: Size 0: 22 x 35 mm Size 1: 24 x 40 mm Size 2: 31 x 41 mm Size 3: 27 x 54 mm Size 4: 57 x 76 mm Similar. The

See the full product information製品情報をすべて見るには