FDA DEVICE FACTS
Reliance® Fusion Sterilization Wraps
Ahlstrom-Munksjo Nonwovens, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
Reliance® Fusion Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:
• Pre-vacuum Steam 270°F/132°C for 4 minutes o Dry time of 12 minutes for Fusion Light o Dry time of 20 minutes for Fusion Heavy o Pre-vacuum steam sterilization of two stainless steel lumens 3 mm in diameter or larger and 400 mm in length or less.
• 100% Ethylene Oxide (EO) with a concentration of 725-759 mg/L @ 131°F/55°C and 40% - 70% relative humidity for 60 minutes o Aeration time of 8 hours at 55oC o Ethylene oxide sterilization of two stainless steel lumens 3 mm in diameter or larger and 400 mm in length or less. Table 1: Reliance® Fusion Models and Load Recommendations Wrap Sheet Size, Maximum Recommended Wrapped Package Content Weight: Pre-Vacuum Steam and Ethylene Oxide1 Wrap Code Material Type Intended Load 12" x 15" x 18" x 24" x 30" x 36" x 40" x 48" x 54" x 54" x 72" 12" 15" 18" 24" 30" 36" 40" 48" 54" Reliance® Fusion Light Outer SMS / Inner Absorbent Wetlaid Layer Light to Moderate Weight Package (e.g. general use medical instruments) 1.8 lb 2.0 lb 2.4 lb 3.2 lb 4.3 lb 5.7 lb 6.7 lb 9 lb 9 lb 9 lb Reliance® Fusion Heavy Outer SMS / Inner Absorbent Wetlaid Layer Very Heavy Weight Package (e.g. general use medical instruments) 1.8 lb 2.6 lb 3.7 lb 6.4 lb 9.9 lb 14.2 lb 17.5 lb 25 lb 25 lb 25 lb 1 It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. In addition, it is recommended to not exceed the number, weight, and size of individual content types that were validated for Reliance® Fusion Sterilization Wraps (i.e. wrapped contents). Table 2: Fusion Wrap Models, sheet color
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。