RARA CRAFT DATABASE for FDA Medical Devices
510(k)K231172

FDA DEVICE FACTS

ApolloKnee

Corin U.S.A. Limited

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJuly 26, 2023PRODUCT CODEOLO

Indications for use

ApolloKnee is indicated to aid the surgeon i n locating anatomical structures and aligning endoprostheses with these anatomical structures during total knee arthroplasty. Apollo Robot and BalanceBot are optional accessories and are specifically i ndicated for use i n total knee arthroplasty with the ApolloKnee system. ApolloKnee is compatible with the following implant systems:

• Apex Knee™ System.

• Unity Knee™ System. ApolloKnee is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate. Summary of Technologies / Substantial Equivalence: The subject ApolloKnee is similar to the OMNIBotics Knee System (per K200888) in terms of design and performance, and the same in intended use. The ApolloKnee is an evolution of the former OMNIBotics platform, including redesigns to modernize the station and software, while maintaining similar function and performance. Furthermore, the ApolloKnee includes additional instrumentation such as the BalanceBot Pre-Resection Paddles, Tibial Cutting Guide and V-array to support new features as in pre-resection balance assessment, robotic assisted tibial resection and resection position validation, respectively. The subject ApolloKnee has been subject to verification and validation testing as described below, with results supporting substantial equivalence with the predicate OMNIBotics Knee System (per K200888). Non-Clinical Testing: Non-clinical testing included:

• Software unit, integration and system testing

• Accuracy testing per custom protocols and ASTM F2554-22

• S ecurity testing

• Electrical safety testing per IEC 60601-1:2005+A1:2012+A2:2020

• Electromagnetic compatibility testing per IEC 60601-1-2:2014+A1:2020

• Various mechanical and functional performance tests per custom protocols Furthermore, surgeon user validation studies (cadaver evaluations) were performed in accordance with IEC 62366-1:2015+A1:2020 and IEC 60601-1-6:2010+A1:2015+A2:2020. The results of the non-clinical testing support substantial equivalence of the subject ApolloKnee and the predicate device.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 18, 2026