RARA CRAFT DATABASE for FDA Medical Devices
510(k)K231895

FDA DEVICE FACTS

Maxtec MaxBlend2+p

Maxtec, LLC

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEDecember 4, 2023PRODUCT CODEBZR

Indications for use

The subject device, Maxtec MaxBlend 2+p,is adding the cleared feature of measuring and monitoring pressure including the disposable accessory lines. This is similar to the secondary predicate K221734 – Maxtec – MaxO2 ME+p.

However, for product classification purposes we have compared to subject device to the primary predicate which has the same primary indications for use with the added pressure monitoring feature. The primary predicate, K161718 is the MaxBlend 2 which is an air/gas blender with oxygen monitoring.

Environment of Use The subject device, Maxtec MaxBlend 2+p has similar environments of use when compared to the primary predicate K161718 Maxtec – MaxBlend 2. There is no difference in the use environments and adding the pressure monitoring feature which has been cleared for the same use environments would raise different concerns of safety or effectiveness.

Population The subject device, Maxtec MaxBlend 2+p has similar patient population as the primary predicate, K161718 Maxtec – MaxBlend 2. While the secondary predicate has been cleared with the pressure monitoring for newborn to adults, the subject device and its primary predicate have been cleared for infants, pediatrics and adults. The addition of the pressure monitoring feature remains within the same patient population as the secondary predicate.

Performance Specifications The subject device, Maxtec MaxBlend 2+p, has similar performance specifications for the blender and oxygen monitoring features as the primary predicate, K161718 – Maxtec MaxBlend 2 and the pressure monitoring performance specifications as the secondary predicate, K221734 – Maxtec MaxO2 ME+p.

Non-clinical Testing We performed a number of tests to demonstrate that the proposed device performed as intended. Testing includes:

• AAMI ANSI ES 60601-1: 2005 +A1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

• IEC 60601-1-2 Edition 4.1 2020 Collateral standard: Electromagnetic Disturbances - Requirements and Tests

• IEC 60601-1-8: 2012 Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems

• ISO 80601-2-55: 2018 Particular requirements for the basic safety and essential performance of respiratory gas monitors

• ISO 11195:2018 Gas mixers for medical use In all cases the subject device met the standard requirements. Biocompatibility The only new patient contacting materials are the disposable pressure tubing which are identical to the secondary predicate, K221734 – Maxtec MaxO2 ME+p. There are no changes in materials or patient contact type. These materials were evaluated in K221734 and included:

• ISO 10993-5:2009 - Cytotoxicity

• ISO 10993-10:2010 – Sensitization and Irritation

• ISO 10993-11:2017 - Material Mediated Pyrogenicity

• ISO 10993-18:2020 – Chemical Characterization with a Toxicological Risk Assessment

• ISO 18562-2:2017 – Particulate Material

• ISO 18562-3:2017 – VOC with Toxicological Risk Assessment

The patient contacting materials were found to be safe for the intended population and indications for use.

Animal No animal testing was performed.

Clinical No human clinical testing was performed.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン