FDA DEVICE FACTS
VUZE System
Vuze Medical , Ltd.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The VUZE System is intended to enable users to load pre-operative 3D images and planning data and register and overlay this data in real time with intra-operative 2D radiographic images of the same anatomy to support device guidance during interventional spinal procedures. The system also offers pre-operative surgical planning including implant sizing, entry location, and trajectory determination along with intra-operative guidance and tool trajectory / position confirmation. Technological Characteristics Comparison:
Substantial Equivalence: The VUZE System is substantially equivalent to the original VUZE System (a.k.a. original device) K210830.
The 510(k) Substantial Equivalence Decision-making Process (detailed) from FDA Guidance - The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] was followed as described below: The modified VUZE System has the same intended use and the same indications for use as the Predicate device (original version of the VUZE System). The VUZE System uses identical fundamental technology as the Predicate device and very similar detailed technological solutions. The following summary of technological changes from the originally cleared device and the modified VUZE System are being made through this 510(k):
o Planning Software: Added option for planning-only station that contains only the planning software which runs on commercial PC meeting certain specified requirements. Option includes export to USB functionality for transferring planning data to VUZE System. Elimination of the need to generate pre-operative data during planning. New feature allows DRR generation in real-time during procedure. Addition of a new feature incorporating horizontal line angle control into the Coronal cross-section, supplementing the existing Sagittal cross-section controls. Add a check of 3D scan slice thickness to ensure it is equal to or greater than slice interval. Discard scans that do not meet this requirement. Add feature that requires the user to indicated whether the skin level is visible in the axial view or not. Add support for loading DICOM-CTs generated by 3D C-arm machines (CBCT). Minor graphical user interface improvements for usability (e.g. view zoom, display simplifications, display pixel spacing, reset button addition, button re-arrangement). Resolution of non-safety related software anomalies.
o VUZE System and Procedure Software Implement a modification to the CT to X-ray registration algorithm to replace the reliance on pre-operative data for generating the initial-guess registration. This algorithm performs in real-time similar calculations that were previously done preoperatively. Implement a registration success score to assess the System’s confidence in the CT to X-ray registration result. This score assesses the System's confidence in the registration and establishes a specific threshold to determine registration failure. Addition of a Bluetooth C-arm orientation sensor and associated Bluetooth dongle to allow compatibility with additional X-ray systems. Deprecation of the previous ML algorithm for determining window size parameters. Same functionality is now done by a traditional algorithm that calculates the filter for several window sizes and chooses the best one. Add option to load CT images and planning data from USB media. Add support for additional C-arm systems for 2D X-ray image acquisition: Philips Zenition 70 GE: OEC Elite CFD Ziehm: Vision FD and Vision RFD 3D (2 different Flat Detector types) Add support for Siemens NXS protocol for image acquisition via a LAN connection as an alternative to using a frame grabber. This option is available only the Siemens Cios Spin C-arms. Add a new “live DRR” display, also known as "Simulated X-ray view" that activates when the C-arm rotates and displays a 2D projection image derived from the 3D CT / CBCT scan data in accordance with the current angles of the C-arm. Add a verification and associated alert to ensure that tool length marking is not shorter than 40mm and that the position of the marked tool is not further than 200 pixels from that marked in the planning data. Add a feature to enable users to instantly view the newly acquired X-ray image prior to its processing, with no dark background. Add a feature that allows the user to verify proper positioning of all the tools (e.g. k-wires prior to screw placement) with two additional AP X-ray images capturing all the tools at once. Add a feature that allows the System to be used in an offline mode so that newly acquired X-ray images aren’t input and analyzed by the VUZE System. Generate cross-sections in the background. If generation fails, the software will suggest acquiring images from different C-arm angles, in advance. If the generation is successful, the user can then view the generated cross-sections within the software by pressing “Generate Cross-Sections”. Addition of a system verification that alerts the user when the tool trajectory changes significantly from what was previously accepted. Addition of an optional Bluetooth foot pedal with associated Bluetooth dongle for the System as an input device that activates the default action from the GUI at the time of activation. Upgrade GPU model from RTX-2080 to RTX-3090 to support faster CT to X-ray registration. Addition of DVI galvanic Isolator and replace existing optical video cable with DVI cable. Minor graphical user interface improvements for usability (e.g., snap to tool, screenshot tool, magnification view, display incision site, C-arm angles display in title bar, always provide next image angle suggestions, text changes). Resolution of non-safety related software anomalies.
The differences in technology do not raise different questions of safety or effectiveness and were evaluated through comprehensive bench verification and validation testing including a usability study. The results of testing provide assurance that the device is as safe and effective as the Predicate. For a more detailed comparison refer to the Substantial Equivalence comparison table included below.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。