ISIS Headboxes, ISIS Neurostimulator, ISIS Xpert Plus, ISIS Xpert, ISIS Xpress
Inomed Medizintechnik GmbH
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
“The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities or their combinations:
• Measurement:
- Auditory evoked potentials (AEP) (…)
• Stimulation:
- Transcranial electrical stimulation (TES)
- Direct cortical and subcortical stimulation (DCS)
- Direct nerve stimulation (DNS)
- Transcutaneous intraoperative nerve stimulation (TINS)
- Direct muscle stimulation (DMS)” - ISIS Headbox 5042XX: “The products support the clinical application of (…) Auditory Evoked Potentials (AEP) (…)” ISIS Neurostimulator 504180: The ISIS Neurostimulator is intended for the provision of neurophysiological stimulation (…). It is suitable for continuous operation and can be used in the following fields:
- Transcranial electrical stimulation (TES)
- Direct cortical stimulation (DCS)
- Direct nerve stimulation (DNS) Equal
AEP: Equal TES: Equal DCS: Equal DNS: Equal TINS and TNS : Equal, the statement was reformulated from transcutaneous electrical nerve stimulation to transcutaneous intraoperative nerve stimulation to highlight the non-therapeutic aspect. DMS: Equal Both the subject and predicate device are indicated for use during surgical procedures that justify the use of aforementioned monitoring modalities. ISIS Headboxes and ISIS Neurostimulator FM 3-1 Stand 1.1 vom 2017-11-02 001_510(k) Summary-2023-10-26.docx
- Transcutaneous electrical nerve stimulation (TNS)
- Direct muscle stimulation (DMS)
“The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities (…)
• Measurement: (…) Electroencephalography (EEG)” ISIS Headbox 5042XX: “The products support the clinical application of Electroencephalography (EEG)” Equal Both the subject and predicate devices are intended for electroencephalography (EEG). “The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities (…)
• Measurement: (…) Electromyography (EMG).” ISIS Headbox 5042XX: “The products support the clinical application of (…) Electromyography (EMG).” Equal Both the subject and predicate devices are intended for electromyography (EMG). “The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities or their combinations:
• Measurement: (…) Motor Evoked Potentials (MEP). (…)
• Stimulation: (…) transcranial electrical stimulation (TES).” ISIS Headbox 5042XX: “The products support the clinical application of (…) Motor Evoked Potentials (MEP).” ISIS Neurostimulator 504180: “It is suitable for continuous operation and can be used in the following fields: (…) transcranial electrical stimulation (TES).” Equal Both devices can deliver transcranial stimulation via dedicated outputs. The ISIS Headbox is capable of recording MEP signals during transcranial electrical stimulation provided by the ISIS Neurostimulator. “The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities (…)
• Stimulation: (…) direct cortical stimulation (DCS).” ISIS Neurostimulator 504180: “It is suitable for continuous operation and can be used in the following fields: (…) direct cortical stimulation (DCS).” Equal Both devices can deliver electrical current for direct cortical stimulation. “The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities or their combinations: ISIS Headbox 5042XX: “The products support the clinical application of (…) Electromyography (EMG).” ISIS Neurostimulator 504180: Equal Both devices can measure triggered EMG signals . The subject devices achieve this by combining the EMG measurement using the ISIS Headboxes and direct nerve ISIS Headboxes and ISIS Neurostimulator FM 3-1 Stand 1.1 vom 2017-11-02 001_510(k) Summary-2023-10-26.docx
• Measurement: (…) Electromyography (EMG).”
• Stimulation: (…) direct nerve stimulation (DNS).” “It is suitable for continuous operation and can be used in the following fields: (…) direct nerve stimulation (DNS).” stimulation (DNS) using the ISIS Neurostimulator. “The product can be used during surgical procedures that justify non-therapeutic clinical use of the following modalities or their combinations: combinations:
• Measurement: (…) Train of Four (TOF)
• Stimulation: (…) Transcutaneous intraoperative nerve stimulation (TINS).” ISIS Headboxes 5042XX: “The products support the clinical application of (…) Electromyography (EMG).” ISIS Neurostimulator 504180: “The ISIS Neurostimulator is intended for the provision of neurophysiological stimulation (…) in the following fields: (…) transcutaneous electrical nerve stimulation (TNS).” Equal Both products can be used for performing twitch tests to measure the level of muscle relaxation due to muscular blockers. This measurement is performed by employing a simultaneous transcutaneous electrical nerve stimulation and EMG measurement. “The products are not intended for monitoring life-sustaining functions.” ISIS Headbox 5042XX: “The products are not intended for monitoring life-sustaining functions.” Equal Both products are not intended for monitoring of life-sustaining function.
Technology
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。