HALYARD® ONE-STEP® Sterilization Wrap, HALYARD® QUICK CHECK® Sterilization Wrap, HALYARD® SEQUENTIAL Sterilization Wrap and HALYARD® SMART-FOLD® Sterilization Wrap
O&M Halyard, Inc.
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
HALYARD® SMART-FOLD Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using: - Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 30 minutes. - 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 131°F/55°C or 12 hours at 110° F/43.3°C. STERIS V-PRO® Low Temperature Sterilization Systems.
• STERIS® V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS® V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS® V-PRO® 1 (Lumen Cycle)
• STERIS® V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)
• STERIS® V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)
• STERIS® V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle) - Advanced Sterilization Products STERRAD® Sterilization System
• STERRAD® 100S
• STERRAD® NX®, (Standard Cycle, Advanced Cycle)
• STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
• STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)
• STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle, ULTRA GI Cycle)
SMART-FOLD Wrap Model Recommendations with Pre-vacuum and EO Halyard SMART-FOLD Sterilization Wrap: H450 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 13 lbs. Halyard SMART-FOLD Sterilization Wrap: H650 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 25 lbs.
SMART-FOLD Wrap Model Recommendations with ASP STERRAD® Sterilization System - ASP STERRAD® 100S, NX®, NX® with ALLClear® Technology, 100NX® and 100NX® with ALLClear® Technology Halyard SMART-FOLD Sterilization Wrap: H450 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 10.7 lbs. Halyard SMART-FOLD Sterilization Wrap: H650 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 10.7 lbs.
SMART-FOLD Wrap Model Recommendations with ASP STERRAD® Sterilization System - ASP STERRAD® 100NX® and 100NX® with ALLClear® Technology - ULTRA GI Cycle Halyard SMART-FOLD Sterilization Wrap: H450 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 7.7 lbs.
Halyard SMART-FOLD Sterilization Wrap: H650 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights***: 7.7 lbs.
SMART-FOLD Wrap Model Recommendations with STERIS V-PRO Sterilizers Halyard SMART-FOLD Sterilization Wrap: H450 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights*** STERIS® V-PRO® 1, 1Plus, maX, maX 2 Lumen: 13 lbs. STERIS® V-PRO® 1Plus, maX, maX 2 Non-Lumen: 13 lbs. STERIS® V-PRO® maX, maX 2 Flexible: 13 lbs. STERIS® V-PRO® 60, s2 Lumen: 11 lbs. STERIS® V-PRO® 60, s2 Non-Lumen: 12 lbs. STERIS® V-PRO® 60, s2 Flexible: The validation studies were conducted with one flexible endoscope packaged into a tray with silicone wrap, and instrument organizers and light cord (if not integral to scope) and additional load.
Halyard SMART-FOLD Sterilization Wrap: H650 Intended Load**: Moderate to Heavyweight Package (for example: general use medical instruments) Wrapped Package Content Weights*** STERIS® V-PRO® 1, 1Plus, maX, maX 2 Lumen: 19.65 lbs. STERIS® V-PRO® 1Plus, maX, maX 2 Non-Lumen: 25 lbs STERIS® V-PRO® maX, maX 2 Flexible: 24 lbs. STERIS® V-PRO® 60, s2 Lumen: 11 lbs. STERIS® V-PRO® 60, s2 Non-Lumen: 12 lbs. STERIS® V-PRO® 60, s2 Flexible: The validation studies were conducted with one flexible endoscope packaged into a tray with silicone wrap, and instrument organizers and light cord (if not integral to scope) and additional load.
** Intended loads include: Medical Instruments with and without lumens that include telescopes, endoscopes, cameras, light cords, and general use medical instruments. *** It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight, and size of individual content types that were validated for the HALYARD* SMART-FOLD* Sterilization Wrap (i.e., the weight of the metal mass).
Validated STERIS® V-PRO® Low Temperature Sterilization Cycles (HALYARD SMART-FOLD) STERIS® V-PRO® 1 and STERIS® V-PRO® 1 Plus Lumen Cycle Intended Load
• Lumened and non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.
• Medical devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations: - Single channeled devices with a stainless lumen that is
• 0.77 mm internal diameter (ID) and 500mm in length
• 1.8 mm ID and 542 mm in length - Dual channeled devices with stainless steel lumens that are 0.77mm ID and 527mm in length - Triple channeled devices with stainless steel lumens that are
• 1.2 mm ID and 275 mm in length
• 1.8 mm ID and 310 mm in length OR
• 2.8 mm ID and 317 mm in length
STERIS® V-PRO® maX and STERIS V-PRO maX 2 Lumen Cycle Intended Load - Medical devices with the following configurations:
• Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.
• Single, dual or triple channeled devices with stainless steel lumens with
• An inside diameter of 0.77 mm or larger and a length of 527 mm or shorter
• An inside diameter of 0.8 mm or larger and a length of 542 mm or shorter
• An inside diameter of 0.48 mm or larger and a length of 100 mm or shorter
- Dead end lumen with
• An inside diameter of 1.3 mm or larger and a length of 73 mm or shorter
-Rigid non-metallic lumen (such as those used in endoscope sheaths, take-apart forceps and trocars) with
• An inside diameter of 3 mm or larger and a length of 298 mm or shorter
• An inside diameter of 4 mm or larger and a length of 424 mm or shorter
Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. The validation studies were performed using a validation load consisting of two trays with silicone mats for a total weight of 19.65 lbs.
STERIS® V-PRO® 60 and s2 Lumen Cycle Intended Load - Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. - Non-lumened devices including non-lumened rigid and semi-rigid endoscopes - Medical Devices, including single, dual and triple channeled rigid and semi-rigid endoscopes, with the following configurations:
• Single or dual lumen devices with stainless steel lumens that are
• 0.77 mm internal diameter (ID) and 410 mm in length
• 1.8 mm ID and 542 mm in length
• Triple lumen devices with stainless steel lumens that are
• 1.2 mm ID and 275 mm in length
• 1.8 mm ID and 310 mm in length or
• 2.8 mm ID and 317 mm in length The Non-Lumen articles can be processed in the Lumen cycle as it takes mixed loads.
STERIS® V-PRO® 1 Plus Non-Lumen Cycle Intended Load Instruments/devices with the following features: - Non-lumened instruments including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened instruments with stainless steel or titanium diffusion-restricted areas such as the hinged portion of forceps or scissors.
STERIS® V-PRO® maX and STERIS® V-PRO® maX 2 Non-Lumen Cycle Intended Load Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes. Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Note: The validation studies (maX and maX 2) were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs.
STERIS® V-PRO® 60 and s2 Non-Lumen Cycle Intended Load Instruments/devices with the following features: - Non-lumened devices including non-lumened rigid, semi-rigid and flexible endoscopes and non-lumened devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors.
STERIS® V-PRO® maX and STERIS® V-PRO® maX 2 Flexible Cycle Intended Load - Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. - Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) or bronchoscopes in either of two load configurations:
• Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load*. The flexible endoscopes may contain single or dual channel lumens that have:
• 1 mm ID and 1050 mm in length Or
• One flexible endoscope with a light cord (if not integral to endoscope), endoscope accessories, mat and additional non-lumened instruments.** The flexible endoscope may contain single or dual channel lumens that have:
• An inside diameter of 1 mm or larger and a length of 1050 mm or shorter Additional instruments may include non-lumened or lumened medical devices with the following configurations:
• Single, dual or triple channel stainless steel lumen that have:
• An inside diameter of 0.48 mm or larger and a length of 100 mm or shorter * The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope). ** The validation studies were conducted with a flexible endoscope in a tray with endoscope accessories, silicone mat, light cord (if not integral to endoscope) and 5 stainless steel lumens. Also included in the load was a tray with additional instruments and silicone mat for a total weight of 24 lbs.
STERIS® V-PRO® 60 and s2 Flexible Cycle Intended Load Load 1:
• One flexible surgical endoscope or bronchoscope with a light cord (if not integral to endoscope), mat, and additional load. The flexible endoscope may be a:
• Single or dual lumen device with lumens that are > 1 mm ID and <990 mm in length
Load 2:*** Non-lumened instruments, instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors, and medical devices containing stainless steel lumens with the following dimensions:
• 0.76 mm ID and 233 mm in length
• 1.0 mm ID and 254 mm in length
• 1.8 mm ID and 542 mm in length *** Additional stainless steel lumens are restricted to 12 lumens Total load weight is restricted to 11 lbs.
Validated Advanced Sterilization Products (ASP) STERRAD® 100S, STERRAD® NX®, STERRAD® NX® with ALLClear® Technology, STERRAD® 100NX® and STERRAD® 100NX® with ALLClear® Technology Cycles
ASP STERRAD® 100S Validation Load Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: - An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens. - An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens. - An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. - An inside diameter of 6 mm or larger and a length of 310 mm or shortcr of single-channel Teflon/Polyethylene lumens. Refer to the STERRAD® 100S Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10 lumens per load).
ASP STERRAD® NX® Standard Cycle Validation Load Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: - An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens. - An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens. Refer to the STERRAD® NX® Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10.7 lbs. per load).
ASP STERRAD® NX® Advanced Cycle Validation Load Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: - An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens, OR One single-channel flexible endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain: - A single-channel Teflon/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. Refer to the STERRAD® NX® Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10.7 lbs. per load).
ASP STERRAD® 100NX® Standard Cycle Validation Load Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: - An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) Refer to the STERRAD® 100NX® Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 21.4 lbs. per load).
ASP STERRAD® 100NX® Flex Cycle Validation Load One or two single-channel flexible endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain: - A single-channel Teflon/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle). Refer to the STERRAD® 100NX® Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 12.2 lbs. per load).
ASP STERRAD® 100NX® EXPRESS Cycle Validation Load Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens. Refer to the STERRAD® 100NX® User’s Guide For complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 10.7 lbs. per load).
ASP STERRAD® 100NX® DUO Cycle Validation Load One or two single-channel flexible endoscopes with accessory devices that are normally connected to it, with or without a silicone mat. The flexible endoscope may contain: - A single-channel Teflon/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 875 mm or shorter. - Accessory devices that are normally connected to a flexible endoscope during use. - Flexible endoscopes without lumens Refer to the STERRAD® 100NX® Sterilizer User's Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 13.2 lbs per load).
ASP STERRAD® 100NX® ULTRA GI Cycle Validation Load
• Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions of 1mm inner diameter x 1500mm in length, or 2mm inner diameter x 1630mm in length.
• One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf. Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes Refer to the STERRAD® 100NX® Sterilizer User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 15.4 lbs. per load).
HALYARD* ONE-STEP, QUICK CHECK and SEQUENTIAL Sterilization Wraps are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:
• Pre-vacuum steam at 270°F/132°C for 4 minutes. The wrap was validated for dry times of 20 minutes for Models 100 and 200, and for 30 minutes for Models 300, 400, 500, and 600.
• 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% relative humidity for 60 minutes. The wrap was validated for aeration times for EO sterilization of 8 hours at 55°C or 12 hours at 43.3°C.
• STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles.
• STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS V-PRO® 1 (Lumen Cycle)
• STERIS V-PRO® 1 Plus (Lumen and Non-Lumen Cycle)
• STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycle)
• STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycle)
• Gravity steam at 250°F/121°C for 30 minutes (25 minute dry time for Models 100, 200 and 300 and 30 minute dry time for Models 400, 500 and 600)
• Advanced Sterilization Products STERRAD® Sterilization System
• STERRAD® 50, 100S, and 200
• STERRAD® NX®, (Standard Cycle, Advanced Cycle)
• STERRAD® NX® with ALLClear® Technology, (Standard Cycle, Advanced Cycle)
• STERRAD® 100NX® (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
• STERRAD® 100NX® with ALLClear® Technology, (Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle)
• STERILUCENT® HC80TT Sterilization Cycles (Lumen and Flexible Cycles)
• Stryker Sterizone® VP4 Sterilizer Cycle 1 HALYARD Wrap Model Recommendations*** Halyard Sterilization Wrap: H100 Intended Load: Very light weight package (e.g., towel packs) Wrapped Package Content Weight****: Pre-vacuum, Gravity and EO: 3 lbs. STERIS V-PRO 1, 1Plus, maX, maX 2: 3 lbs. STERIS V-PRO 60, s2: 3 lbs. ASP STERRAD 50, 100S, 200, NX®, NX® with ALLClear® Technology, 100NX® and 100NX® with ALLClear® Technology Cycles: 10.7 lbs. STERILUCENT HC80TT Lumen and Flexible Cycles: 3.4 lbs. Stryker Sterizone VP4 Sterilizer Cycle 1: 3 lbs.
Halyard Sterilization Wrap: H200 Intended Load: Light weight package (e.g., standard linen packs, telescope with light cord) Wrapped Package Content Weight****: Pre-vacuum, Gravity and EO: 6 lbs. STERIS V-PRO 1, 1Plus, maX, maX 2: 6.5 lbs.
STERIS V-PRO 60, s2: 6.5 lbs. ASP STERRAD 50, 100S, 200, NX®, NX® with ALLClear® Technology, 100NX® and 100NX® with ALLClear® Technology Cycles: 10.7 lbs. STERILUCENT HC80TT Lumen and Flexible Cycles: 6 lbs. Stryker Sterizone VP4 Sterilizer Cycle 1: 6 lbs.
Halyard Sterilization Wrap: H300 Intended Load: Light to moderate weight package (e.g., general use medical instruments) Wrapped Package Content Weight****: Pre-vacuum, Gravity and EO: 9 lbs. STERIS V-PRO 1, 1Plus, maX, maX 2: 9 lbs. STERIS V-PRO 60, s2: 9 lbs. ASP STERRAD 50, 100S, 200, NX®, NX® with ALLClear® Technology, 100NX® and 100NX® with ALLClear® Technology Cycles: 10.7 lbs. STERILUCENT HC80TT Lumen and Flexible Cycles: 9.1 lbs. Stryker Sterizone VP4 Sterilizer Cycle 1: 9 lbs.
Halyard Sterilization Wrap: H400 Intended Load: Moderate to heavy weight package (e.g., general use medical instruments) Wrapped Package Content Weight****: Pre-vacuum, Gravity and EO: 13 lbs.***** STERIS V-PRO 1, 1Plus, maX, maX 2: 10 lbs. STERIS V-PRO 60, s2: 12 lbs. ASP STERRAD 50, 100S, 200, NX®, NX® with ALLClear® Technology, 100NX® and 100NX® with ALLClear® Technology Cycles: 10.7 lbs. STERILUCENT HC80TT Lumen and Flexible Cycles: 13 lbs. Stryker Sterizone VP4 Sterilizer Cycle 1: 13 lbs.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。