RARA CRAFT DATABASE for FDA Medical Devices
510(k)K240438

FDA DEVICE FACTS

CDB Hospital ® Sterilization Wrap (70 gsm); CDB Hospital ® Sterilization Wrap (44 gsm)

Comercializadora Brever

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATENovember 4, 2024PRODUCT CODEFRG

Indications for use

CDB Hospital ® Sterilization Wraps intended to be used to enclose another medical device that is to be sterilized by a healthcare provider via the following:

• Pre-vacuum Steam 270°F/132°C for 4 minutes

• Advanced Sterilization Products (ASP) STERRAD® 100S

• Advanced Sterilization Products (ASP) STERRAD® 100NX (Standard, Express and Flex cycles)

• STERIS Amsco® V-PRO 1, STERIS Amsco® V-PRO 1 Plus, STERIS Amsco® V-PRO maX Low Temperature Sterilization Systems (Lumen, Non Lumen and Flexible Cycles) The wrap is intended to allow sterilization of the enclosed medical device(s) and maintain sterility of the enclosed device(s) until used. Pre-Vacuum Steam Sterilization:

• Validated for dry time of 30 minutes for 44gsm.

• Validated for dry time of 30 minutes for 70gsm.

• Validated maximum weight of 6 lbs for 44 gsm.

• Validated maximum weight of 15 lbs for 70 gsm.

• All models of CDB Hospital ® Sterilization Wraps validated for pre-vacuum steam sterilization with stainless steel lumens of 3 mm diameter or larger and length 400 mm or shorter as part of the load. Advanced Sterilization Products (ASP) STERRAD® 100S Sterilization: All models of CDB Hospital® Sterilization Wraps validated with ASP STERRAD® 100S sterilization with stainless steel lumens as part of the load with a maximum weight of 10.7lbs. Advanced Sterilization Products (ASP) STERRAD® 100NX Sterilization: All models of CDB Hospital® Sterilization Wraps validated with ASP STERRAD® 100NX sterilization cycles are detailed below: VALIDATED ADVANCED STERILIZATION PRODUCTS (ASP) STERRAD® 100NX CYCLES Advanced Sterilization Products (ASP) STERRAD® System and Cycle Intended Load for 100NX Standard Cycle: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:

• An inside diameter of 0.7mm or larger and a length of 200mm or shorter of single channel stainless steel lumens. Intended Load for 100NX Flex Cycle: One or two single-channel flexible endoscopes with or without a silicone mat. Flexible endoscope may contain: A single channel Teflon/PE/PTFE lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter (a maximum of two flexible endoscopes, one per tray per sterilization cycle). Intended Load for 100NX Express Cycle: Non-lumened reusable metal and non-metal medical devices requiring surface sterilization, or sterilization of mated stainless steel and titanium surfaces, and rigid or semi-rigid endoscopes without lumens. STERIS Amsco® V-PRO 1, STERIS Amsco® V-PRO 1 Plus, STERIS Amsco® V-PRO maX Low Temperature Sterilization Systems:

• All models of CDB Hospital® Sterilization Wraps have been validated for use with STERIS V-PRO® cycles detailed below.

• CDB Hospital® Sterilization Wraps were validated to be effectively aerated during the preprogrammed STERIS Amsco V-PRO® Sterilization Cycles. VALIDATED STERIS AMSCO® V-PRO CYCLES STERIS Amsco® V-PRO Cycle Intended Load for Lumen Cycle: Reusable metal and non-metal medical devices, including up to 20 stainless steel lumens with dimensions of 3.0mm diameter or larger and a length of 400mm or shorter. Intended Load for Non Lumen Cycle: Non lumened reusable metal and non-metal medical devices. Intended Load for Flexible Cycle: Single lumen surgical flexible endoscopes and bronchoscopes in the following load configuration: 1.) One tray containing a flexible endoscope with a light cord (if not integral to endoscope) and mat. The flexible endoscope having an inside diameter of 3 mm or larger and a length of 450 mm or shorter. 2.) An additional tray containing non-lumened medical devices.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 16, 2026