RARA CRAFT DATABASE for FDA Medical Devices
510(k)K242132

FDA DEVICE FACTS

FREDDO

Fabinject Lndustria E Comercio Lmportacao E Exportacao Ltda

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEOctober 18, 2024PRODUCT CODEGEX

Indications for use

The Freddo is intended to minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections. The Cryo V6.0 Cold Air Device is intended to minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.

Identical Performance Characteristics Measurement at device outlet -17°C to -28°C (± 3°C) At an air speed of 160 to 780 liters/minute (± 10%). -31°C At an air speed of 100 to 1000 liters/minute. Similar, not significantly different

Cooling Method, Operation principle Cooling agent-based condenser and evaporation system. Cold air is produced which is then directed onto the skin. Conduction, evaporation, and forced convection cool the skin. Cooling agent-based condenser and evaporation system. Cold air is produced which is then directed onto the skin. Conduction, evaporation, and forced convection cool the skin.

Identical Cooling Material Gas (Air) Gas (Air) Identical Mains Voltage 120V / 60Hz 100V – 120V / 50-60Hz Similar, not important for the device safety and effectiveness evaluation Main Fuse 16A circuit breaker in main switch for 120V 10A circuit breaker in main switch for 220V 16A circuit breaker in main switch Similar, not important for the device safety and effectiveness evaluation Power Consumption Max. 11A Max. 9 - 11A No impact IP classification IPX0 IPX0 Identical Design Metal enclosure Metal enclosure Identical Materials Environmentally approved R404, closed loop cooling system Environmentally approved R507, closed loop cooling system Both environmentally approved, no impact Cooling spot Size Varies with distance from distal end of air hose to tissue – 10cm² @ 5cm distance Varies with distance from distal end of air hose to tissue – 10cm² @ 5cm distance Identical Fan speed settings 1 - 3 1 - 9 No impact Comparison with the

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: August 16, 2026