PAL Sterilization Case
Microaire Surgical Instruments, LLC
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The PAL Sterilization Ca se is a reusable ste rilization case intended to provide sto rage for the PA L handpiece, the P AL cable, and 4 PAL cannulas during sterilization, sto rage, and transportatio n within the hospital environment. The PAL Steri liz ation Case is only intended to ma intain steri lity of the enclosed components during auto clave sterilizati on when used in conjunction with FD A-cleared steril izatio n wraps. The PA L Steriliz ation Case has only been evaluated for a nonstacked configuration. The validate d steam sterilizati on param eters are listed below: Cycle Type Pre-Vacuum Gravity Exposure Tem perature 132 °C (270 °F) 121 °C (250 °F) Exposure Time (m inutes) 4 m inutes 30 minutes M inim um Dry Time (m inutes) 20 minutes 20 minutes Com patible Sterile Barr iers W rap(s) W rap(s)