Chex-All Sterilization Pouches and Tubes
Propper Manufacturing Co., Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
Indications for use: Chex-All® Sterilization Pouches and Tubes are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using: Hospi tal size sterilizer cycles
•Gravit y steam at 121°C (250°F) for 30 minutes; 25 minutes dry time
• Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
• Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time Tabletop steriliz er cycles:
• Pre-vacuum steam at 132°C (270°F) for 4 minutes; 30 minutes dry time Ethylene Oxid e sterilization cycles:
•Ethylene Oxide (EO) with a concentration of 736 mg/L at 55°C (131°F) and 50% to 60% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). Chex-All sterilization pouches and tubes can be used for steam sterilization of devices with lumens at maximum load as the following: Metal lumen: 7 inch long, 3.2mm internal diameter Plastic lumen: 5 inch long, 2.0 mm internal diameter. Suitable Lumen cycles: Hospital size sterilizers: pre-vacuum cycles with 132°C-4.0 min exposure (20 minutes dry time) and 135°C-3.0 min exposure (16 minutes dry time). Table-top sterilizers: pre-vacuum cycle with 132°C-4.0 min exposure (30 minutes dry time). Chex-All sterilization pouches and tubes can be used for steam sterilization in double pouch configuration, in gravity 250°F-30 min exposure cycles (25 minutes dry time) and in pre-vacuum 270°F-4 min exposure cycles (20 minutes dry time). The Process chemical indicators on the Chex-All sterilization pouches and tubes are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed and unprocessed pouches. The chemical indicators change from crème to gray/black after exposure to steam and from beige-pink to green after exposure to ethylene oxide. If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 2 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and to maintain sterility (SAL≥10-6). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 30 months after steam sterilization and 12 months after EO sterilization. The maximum pouch load is 2.65 lb. (1.2 kg) or less. The maximum pouch load for table-top sterilizers is 1.10 lb. (0.5 kg) or less. The maximum pouch size for use in table-top sterilizers is 13”x18”. Cat. Numbers External Dimensions Max load weight Chex-All Heat-seal Width Length lb sterilization pouch, flat 02500800 3” 8” 0.99 02501000 5” 10” 0.99 02501200 7” 12” 1.10 02501400 8” 16” 1.10 02501800 12” 18” 2.65 02502400 18” 24” 2.65 Chex-All Heat-seal sterilization tube, flat, sold in rolls 02600200 2” 100’ 0.99 02600300 3” 100’ 0.99 02600400 4” 100’ 0.99 02600600 6” 100’ 1.10 02600900 9” 100’ 1.10 02601200 12” 100’ 2.65 Chex-All II Self-seal sterilization pouch, flat 02400800 3” 8” 0.99 02401000 5” 10” 0.99 02401200 7” 12” 1.10 02401400 8” 16” 1.10 02401500 5” 15” 0.99 02401800 12” 18” 2.65 02402800 3.5” 24” 0.99 Chex-All III Self-seal sterilization pouch, flat 02430900 3.5” 9.5” 0.99 02431000 5.25” 10” 0.99 02431300 7.5” 12” 1.10 02431400 9” 16” 1.10 02431800 13” 18” 2.65
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。