RARA CRAFT DATABASE for FDA Medical Devices
510(k)K250321

FDA DEVICE FACTS

GCI Sterilization Wrappers

George Courey, Inc.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATEJune 6, 2025PRODUCT CODEFRG

Indications for use

GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam

sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days.

GCI Sterilization Wrapper sizes and validated load limitations are as follows:

SKU / Model Size Border Color Maximum Validated Load Weight

P81818 KBRO 18”x18” White 1 Ibs

P82424 KBRO | 24”x24” Pink 1.82 Ibs

P83030 KBRO 30”x30” Yellow 3.45 Ibs

P83636 KBRO 36”x36” Blue 5.8 Ibs

P84545 KBRO_ | 45”x45” Mint 10.98 Ibs

P85454 KBRO 54”x54” Red 18.5 Ibs

P86060 KBRO | 60”x60” Green 25 Ibs

GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures.

Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: September 1, 2026