DECISION DATEJuly 25, 2025PRODUCT CODELPH

Indications for use

Restoration ADM and MDM Systems (ADM/MDM X3 Inserts, MDM Acetabular Inserts, MDM Acetabular Liners) Indications for Use: The indications for use for total hip arthroplasty include:

1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;

2. Rheumatoid arthritis;

3. Correction of functional deformity;

4. Revision procedures where other treatments or devices have failed; and,

5. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

6. Dislocation risks MDM Liners are intended for cementless use only. Trident II Acetabular System (Trident II Cups (Clusterhole HA, PLS Clusterhold HA, Tritanium Clusterhole, Tritanium Multihole, Tritanium Solidback) Indications for Use:

• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. When used with MDM Liners

• Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

• Dislocation risks When used with Constrained Liner:

• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability. The Trident II Acetabular Shells are indicated for cementless use only. Restoration® Modular Hip System The Restoration® Modular Hip System is indicated for use in:

• Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;

• Rheumatoid arthritis;

• Correction of functional deformity;

• Revision procedures where other treatments or devices have failed; and,

• Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indications specific to the Restoration Modular Hip System The Restoration® Modular Hip System is intended to be used for primary and revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur. Secur-Fit Advanced The indications for use for total hip arthroplasty with stems include:

• noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;

• rheumatoid arthritis;

• correction of functional deformity;

• revision procedures where other treatments or devices have failed; and,

• nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:

• When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability. The Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures. Accolade II Femoral Stems The indications for use of the total hip replacement prostheses include:

• noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;

• rheumatoid arthritis;

• correction of functional deformity;

• revision procedures where other treatments or devices have failed; and,

• nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indication specific to use of the Femoral Stems with compatible Howmedica Osteonics Constrained Liners:

• When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra- operative instability. The Stems are intended for cementless use only and are intended for total and hemiarthroplasty. TRIDENT AND TRITANIUM ACETABULAR COMPONENTS

• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. The HOWMEDICA OSTEONICS TRIDENT and TRITANIUM Acetabular Shells are intended for cementless use only. Dome hole plug is indicated for cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous V40 COBALT-CHROME (COCR) FEMORAL HEADS (LFIT V40™ Femoral Heads)

• Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

• Revision of previous cup arthroplasty or other procedures

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum. Insignia Hip Stem Hip Arthroplasty Indications:

• Painful, disabling joint disease of the hip resulting from: noninflammatory degenerative joint disease (including osteoarthritis or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.

• Revision of previous unsuccessful femoral head replacement, hip arthroplasty or other procedure.

• Correction of functional deformity

• Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indication specific to use of lnsignia Hip Stems with compatible Howmedica Osteonics Constrained Liners:

• When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability. Additional indication specific to use of lnsignia Hip Stems with compatible ADM and MDM Acetabular Components:

• When the stem is to be used with compatible Howmedica Osteonics ADM and MDM Acetabular Components, the device is indicated for Dislocation risks Insignia Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures. Restoration Anatomic Shells

• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.

• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. When used with MDM Liners

• Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

• Dislocation risks When used with Constrained Liner:

• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability. The Restoration® Anatomic Shell is indicated for cementless use only. Trident X3, Trident X3 Elevated Rim, Trident X3 Eccentric Inserts) Indications for Use:

• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum. BIOLOX Delta Ceramic V40 Femoral Heads (V40) The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures. For Use as a Total Hip Replacement:

• Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.

• Revision of previous cup arthroplasty or other procedures.

• Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.

• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum. For Use as a Bipolar Hip Replacement

• Femoral head/neck fractures or non-unions.

• Aseptic necrosis of the femoral head.

• Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.

• Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.

• Salvage of failed total hip arthroplasty “D” femoral heads

• The Novae® Dual Mobility and Hype® hip Systems are indicated for total hip replacement, which includes:

- osteoarthritis,

- femoral neck fracture,

- dislocation risk,

- osteonecrosis of the femoral head,

- revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

• Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use.

• Hype® SCS, SCC, SCL, SCV, SCC Mini, and SCLA Mini hip stems are intended for press-fit use. “C” femoral heads The Novae® Dual Mobility and Hype® hip Systems are indicated for total hip replacement, which includes:

- osteoarthritis,

- femoral neck fracture,

- dislocation risk,

- osteonecrosis of the femoral head,

- revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

• Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use.

• Hype® SCS, SCC, SCL, SCV, SCC Mini, and SCLA Mini hip stems are intended for press-fit use. Hype® SC cementless hip stems The Novae® Dual Mobility and Hype® hip Systems are indicated for total hip replacement, which includes:

- osteoarthritis,

- femoral neck fracture,

- dislocation risk,

- osteonecrosis of the femoral head,

- revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

• Sunfit TH, Novae® E TH, and Coptos TH acetabular cups are intended for press-fit use and Novae® Stick acetabular cup is indicated for cemented use.

• Hype® SCS, SCC, SCL, SCV, SCC Mini, and SCLA Mini hip stems are intended for press-fit use. Novae® dual mobility acetabular cup range including Novae® cementless or cemented metal‑backs, CI E and XPEO‑E liners, pegs and VCI screws

• Novae® dual mobility acetabular cups are indicated for total hip replacement, which includes:

- osteoarthritis,

- femoral neck fracture,

- dislocation risk,

- osteonecrosis of the femoral head,

- revision procedures where other treatments or devices have failed and if bone reconstruction so permits.

• “Sunfit TH”, “Novae E TH” and “Coptos TH” acetabular cups are intended for press-fit use.

• Novae® Stick acetabular cup is intended for cemented use.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: September 1, 2026