Telemetry Monitor (iT30, iT50)
Edan Instruments, Inc.
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
The iT30&iT50 telemetry monitor can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of displaying, storing, and transferring of multiple physiological parameters connecting to Central Monitoring System. It will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient’s situation as appropriate. The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients (over three years old) to monitor physiological parameters include ECG , SpO2, PR, NIBP (Applicable to iT50 only), and Skin TEMP . The physiological data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can configure, display and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities.
| Thermometer, Electronic, Clinical | |
|---|---|
| 21 CFR 870.2700 Oximeter | DQA |
| Characteristic | <Subject Device> (iT30, iT50 Telemetry Monitor) | <Predicate Device> (TM80 Telemetry Monitor) |
|---|---|---|
| Manufacturer/K# | K251508 | K193391 |
| Description | The iT30 and iT50 Telemetry Monitor are intended for use on Adult and Pediatric patients | The TM80 telemetry monitor is intended for use on Adult and Pediatric patients over three |
| (over three years old) to monitor physiological parameters include ECG, SpO2, PR, NIBP (Applicable to iT50 only), and Skin physiological TEMP. The data can be analyzed, alarmed, stored locally on the monitor, and the Edan Central Monitoring System can display configure, and review the physiological parameters from the iT30 and iT50 Telemetry Monitor. The arrhythmia detection and ST Segment analysis of ECG are intended for adult patients. The SpO2 and PR are used under no motion conditions. The NIBP is used under no motion conditions for patients over 12 years old and excluding pregnant women. The iT30 and iT50 Telemetry Monitor must be used by professional trained clinical medical staff in healthcare facilities. | years old to monitor ECG, SpO2, NIBP and Resp physiological data. The physiological data can be analyzed, alarmed, stored, reviewed locally on the display of the monitor, and the CentralStation can configure and display the physiological parameters from the TM80. The TM80 telemetry monitor is intended for use in professional healthcare facilities under the direct supervision of a licensed healthcare practitioner. | |
| Lead Mode | 3 Electrodes; 5 Electrodes; 6 Electrodes; 10 Electrodes | 3 Electrodes; 5 Electrodes; 6 Electrodes |
| Measurement Range | -2.0 mV to +2.0 mV | -2.0 mV to +2.0 mV |
| Pulse Indicator | Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to2.0 ms Ascending time: 10 μs to 100 μs | Amplitude: ±2 mV to ±700 mV Width: 0.1 ms to 2 ms Rise time: 10 μs to 100 μs |
| Measurement range | ADU: 15 bpm to 300 bpm PED: 15 bpm to 350 bpm | Adult: 15 bpm to 300 bpm Pediatric: 15 bpm to 350 bpm |
| QT/QTc/ΔQTc measurement | QT Range: 200 ms ~ 800 ms QTc Range :200ms ~ 800 ms ΔQTc Range: -600 ms ~ 600 ms | QT Range: 200 ms ~ 800 ms QTc Range :200ms ~ 800 ms ΔQTc Range: -600 ms ~ 600 ms |
| Arrhythmia analysis | ASYSTOLE, VFIB/VTAC, COUPLET, VT > 2, BIGEMINY, TRIGEMINY, VENT, R on T, PVC, TACHY, BRADY, MISSED | Asystole, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme Brady, PVCs/min, Pauses/min, R on T, Run PVCs, |
| BEATS, IRR, VBRADY, PNC, PNP | Couplet, Multiform PVC, PVC, Bigeminy, Trigeminy, Tachy, Brady, Pacer Not Pacing, Pacer Not Capture, Missed Beats, Nonsus V-Tach, Vent Rhythm, Pause, Irr Rhythm, A-Fib | |
| Technique | Oscillometry | Oscillometry |
| Measurement Mode | Manual, Auto, STAT | Manual, Automatic, Continuous, Sequence and ABPM |
| Measurement Algorithm | iCUFS | The method principle is equivalent to the iCUFS algorithm |
| Measurement Range | Adult SYS: 25 mmHg to 290 mmHg DIA: 10 mmHg to 250 mmHg MAP: 15 mmHg to 260 mmHg Pediatric SYS: 25 mmHg to 240 mmHg DIA: 10 mmHg to 200 mmHg MAP: 15 mmHg to 215 mmHg | Adult SYS: 25 mmHg to 290 mmHg DIA: 10 mmHg to 250 mmHg MAP: 15 mmHg to 260 mmHg Pediatric SYS: 25 mmHg to 240 mmHg DIA: 10 mmHg to 200 mmHg MAP: 15 mmHg to 215 mmHg |
| Measurement range | 40 bpm to 240 bpm | 30 bpm to 300 bpm |
| Measurement Range | 0% to 100% | Nellcor SpO2 0% to 100% |
| Accuracy | Adult/pediatric, non-motion conditions 2% (70% to 100% SpO2) Undefined (0% to 69% SpO2) | Nellcor SpO2 70-100%: ±2% 0-69%: Not specified |
| Measurement Range | 20 bpm to 300 bpm | 20 bpm to 300 bpm |
| Accuracy | 2 bpm | 20 bpm to 250 bpm: ±3 bpm 251 bpm to 300 bpm: Not specified |
| Number of channels | 1 | |
| Measurement Range | 0 C to 50 C(32 ºF to 122 ºF) | |
| Measurement Site | Skin | |
| Radio Technology | IEEE 802.11 a/b/g/n | IEEE 802.11 a/b/g/n/ac |
| Frequency Range | 2412 MHz-2472 MHz | 2412 MHz to 2472 MHz |