RARA CRAFT DATABASE for FDA Medical Devices
510(k)K251756

FDA DEVICE FACTS

Sterilization Trays

Paragon Implant Mfg., LLC

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATENovember 14, 2025PRODUCT CODEKCT

Indications for use

The Sterilization Trays are designed to hold various dental surgical and prosthetic instruments to organize, steam sterilize and transport the instruments between uses. The kit is to be enclosed in an FDA cleared steam sterilizable pouch large enough to fit a tray, and sterilized in an FDA cleared sterilizer for the following cycles:

1. Gravity Steam Cycle — At 132°C for 15 minutes with a 30-minute dry time

2. Pre-vacuum Steam Cycle — At 132°C for 4 minutes with a 20-minute dry time

* The kits are intended for sterilization of non-porous loads * The tested kit represents the worst-case validated load of 497.1 grams

Model Model Max # of Total Mass Vent to

Name (Catalog) Instruments | (g) Volume Number Ratio (cm-

1)

Complete CSK 55 497.1 0.01043

Surgical Kit

Standard SSK N/A 372.1 0.01043

Surgical Kit

Drill Stop DSK 20 37.5 0.9009

Kit

Drill Stop Not N/A 31.2 0.9009

Kit Empty available

Prosthetic PKE N/A 120.3 0.03324

Kit Empty

Prosthetic Not 20 497.1 0.03324

Kit available

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: September 1, 2026