FDA DEVICE FACTS
Ultima System (UL-800);Ultimate Face Package (UL-800-Face-Set);Ultimate Shape Extension Kit 1 (UL-800-Shape-Ext-Kit-1);Ultimate Shape Extension Kit 2 (UL-800-Shape-Ext-Kit-2)
Fit-Pro KFT (Ltd.) - DBA E-Fit
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
The ULTIMA UL-800 System is intended for use in adults 22 years of age and older. The system is designated for for prescription use only in professional healthcare settings.
The ULTIMA UL-800 System provides two energy modalities, electrical stimulation and light output. Electrical stimulation has separate body NGX and facial NFO applications. Electrical stimulation and light output are operated separately and cannot be delivered simultaneously.
NGX electrical muscle stimulation function The ULTIMA UL-800 System is intended to apply electrical current through surface electrodes to healthy muscles for muscle-conditioning purposes by temporarily affecting the stimulated muscle’s contractile properties, force output, and/or fatigue resistance. The NGX function is not intended for rehabilitation or for the treatment of patients with medical conditions. NGX is available with UL-PAD-IR1 / IR2 / IR3 / IR4 for the labeled body parts.
NFO facial aesthetic electrical stimulation function When UL-FACE is operated in the separate NFO electrical stimulation mode, the ULTIMA UL-800 System is indicated for aesthetic use including facial skin stimulation. Application is limited to the labeled forehead and cheek sites on intact skin. This function is not intended for rehabilitation, treatment of medical conditions, or conditioning of facial muscles to improve contractile properties, force output, or fatigue resistance.
ILY topical-heating light function
| Device & Predicate Device(s): | K260945 | K202148 | K242532 |
|---|---|---|---|
| General Device Characteristics | Subject Device | Primary Predicate | Secondary Predicate |
| Trade Name | Ultima System (ULTIMA UL-800) | E-Fit Infinity System (EF-100, EF-2090) | BTL-785BNF-E |
| Manufacturer | Fit-Pro KFT, Ltd. | Fit-Pro Ltd. | BTL Industries, Inc. |
| Reg. Name | Powered muscle stimulator, for muscle conditioning; Infrared therapeutic heating lamp | Powered muscle stimulator, for muscle conditioning | Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Pro. Code | NGX, NFO, ILY | NGX | NFO |
| Reg. Number | 890.5850, 882.5890, 890.5500 | 890.5850 | 882.5890 |
| Class | II | II | II |
| Intended Use | The ULTIMA UL-800 System is intended for use in adults 22 years of age and older. The system is designated for for prescription use only in professional healthcare settings. The ULTIMA UL-800 System provides two energy modalities, electrical stimulation and light output. Electrical stimulation has separate body NGX and facial NFO applications. Electrical stimulation and light output are operated separately and cannot be delivered simultaneously. NGX electrical muscle stimulation function The ULTIMA UL-800 System is intended to apply electrical current through surface electrodes to healthy muscles for muscle-conditioning purposes by temporarily affecting the stimulated muscle’s contractile properties, force output, and/or fatigue resistance. The NGX function is not intended for rehabilitation or for the treatment of patients with medical conditions. NGX is available with UL-PAD-IR1 / IR2 / IR3 / IR4 for the labeled body parts. NFO facial aesthetic electrical stimulation function When UL-FACE is operated in the separate NFO electrical stimulation mode, the ULTIMA UL-800 System is indicated for aesthetic use including facial skin stimulation. Application is limited to the labeled forehead and cheek sites on intact skin. This function is not intended for rehabilitation, treatment of medical conditions, or conditioning of facial muscles to improve contractile properties, force output, or fatigue resistance. ILY topical-heating light function | E-Fit Infinity System (EF-2090 and EF-100) is a machine system with electronic muscle stimulation based on EMS technology. Regarding its use, the output power is limited and specifically designed as an addition to other sports and for toning muscles. It may only be used for healthy muscles and clients, and not to be used for rehabilitation purposes. The E-Fit Infinity system was designed to stimulate healthy muscles in order to improve or facilitate muscle performance. The E-Fit Infinity system must not be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the E Fit Infinity system training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. The E-Fit Infinity system's electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the E-Fit Infinity system can impose on the stimulated muscles are able to improve or facilitate muscle performance. The E-Fit Infinity system may therefore be considered a technique of muscle training. | The device is indicated for aesthetic use including facial and neck or body skin stimulation. |
| General Device Characteristics | Subject Device | Primary Predicate | Secondary Predicate |
| Trade Name | Ultima System (ULTIMA UL-800) | E-Fit Infinity System (EF-100, EF-2090) | BTL-785BNF-E |
| When UL-FACE is operated in the separate ILY light mode, the ULTIMA UL-800 System is intended to emit energy in the visible red and near-infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, temporarily increasing local blood circulation, and temporarily relaxing muscles. | |||
| Intended Users | Trained Healthcare Professionals | Trained Healthcare Professionals | Trained Healthcare Professionals |
| Use Environment | Professional Healthcare Facility | Professional Healthcare Facility | Prescription Use Setting |
| User Interface | The wireless controller has rotary encoder which allows for a quick set-up, and because of the push button capability, the program can be stopped immediately for every channel. There are large START/STOP and POWER off buttons to begin the program and for complete power shutdown. Because of the pictographs on the displays and fixed electrodes in the clothing, it is very easy to set the appropriate muscle groups. | Main control unit with touchscreen controller. Stimulation is provided by a handpiece with single-use application electrodes connected via cables to the main console. | |
| Prescription | Prescription | Prescription | Prescription |
| Power Source | External AC/DC Power supply; 24 V DC; 220 W | Battery 3.7 V Lithium Polymer (x2) | Mains Power, 100-240 V, 50 – 60 Hz |
| Treatment Time | 30 min | 30 min | Up to 20 min |
| Output Modes | Two Electrical (EMS, TENS); Infrared Light Stimulation provided independent of each other. | One EMS | One TENS |
| Auto-shutoff | Startup self-test, current/voltage limits, no-load/contact detection, watchdog, hardware Health Protection/relay isolation, DC blocking, safe OFF on stop/fault/reset/interruption, and no automatic restart. | Patient discomfort shut-off button | |
| Electrode Materials | 316L Stainless Steel | 316L Stainless Steel | Conductive Silver ink |
| Patient Electrode Interface | Reusable stainless-steel electrode through patient-assigned UL-UC textile (96% cotton / 4% elastane), uniformly moistened only with potable water. | Electrode through 100% hygroscopic cotton moistened with tap water; no direct electrode-to-skin placement. | Electrode via electroconductive hydrogel. |
| Anatomical Location of Stimulation | Face (TENS): 1. M. Frontalis 2. M. Buccinator 3. M. Masseter | 1. Chest 2. Arms 3. Abdominals 4. Back | Face, neck and body |
| General Device Characteristics | Subject Device | Primary Predicate | Secondary Predicate |
| Trade Name | Ultima System (ULTIMA UL-800) | E-Fit Infinity System (EF-100, EF-2090) | BTL-785BNF-E |
| Body (EMS): 1. Abdominis 2. Rectus Femoris 3. Vastus Medialis 4. Vastus Lateralis 5. Gluteus Maximus 6. Musculi Ischiocrurales | 5. Thighs 6. Buttocks 7. Glutes 8. Hamstrings 9. Calves | ||
| Reusable | Yes | Yes | No, Single Use electrodes |
| Electrode Size (cm2) | Face: 45.21 (aggregate); Body: 114.7 (per applicator) | Not available. | 1: 3.8 cm² 2: 4.0 cm² 7: 1.9 cm² 8: 2.9 cm² 9: 3.5 cm² 11: 1.8 cm² |
| Number of Channels | 1 output channel, Time-separated switching to 11 outputs/channels; and electrical current can be regulated individually on every output. | 1 output channel can shift in time between the 12 outputs and electrical current can be regulated individually on every output. | Not publicly available |
| Max Output Current (@ 500 Ω) | 75 mA | 75 mA | G1: 97 mA G2: 64 mA G3: 30 mA |
| Max Output Voltage (@ 500 Ω) | 37.5 V | 37.2 V | 48.5 V |
| Max Phase Charge µC (@ 500 Ω) | Face: 13; Body: 24 | 26 | G1: 15 μC G2: 10 μC G3: 5 μC |
| Max Current Density (mA/cm2) (@ 500 Ω) | 0.6538 | 0.9365 | 1: 16.8; 2: 24.3; 7: 33.7; 8: 33.5; 9: 27.7 11: 16.4 mA/cm² Not identified as RMS. |
| Max Power Density (W/cm2) (@ 500 Ω) | 0.0306 | 0.0342 | 1: 0.033; 2: 0.047; 7: 0.065; 8: 0.065; 9: 0.054; 11: 0.032 |
| Waveform | Charge-balanced biphasic rectangular pulse modulated by trapezoidal | Symmetric biphasic | Charge-balanced biphasic rectangular pulse modulated by trapezoidal |
| Frequency (Hz) | 4 - 220 | 4 - 90 | 250 |
| Pulse Duration | 100 - 500 µs | 100 - 500 µs | 160 µs |
When UL-FACE is operated in the separate ILY light mode, the ULTIMA UL-800 System is intended to emit energy in the visible red and near-infrared spectrum to provide topical heating for the purpose of elevating tissue temperature, temporarily increasing local blood circulation, and temporarily relaxing muscles.
Tables marked from the FDA document were reconstructed automatically. Check the original FDA document (links at the end of this page) for authoritative values.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。