Trevo Trak 21 Microcatheter
Stryker Neurovascular
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Microcatheter is indicated to provide vascular access for selective placement of devices and/or fluids, such as contrast media, within the neurovasculature, including large intracranial vessels (e.g., internal carotid artery [ICA] and middle cerebral artery [MCA]), during diagnostic and/or therapeutic procedures. Intended Use/Indication for Use The Microcatheter is indicated for use in the selective placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures. Device Description The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coatings on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. The Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque marker(s) on the distal end and a luer hub on the proximal end. The catheter shaft has a hydrophilic coating on the distal 100cm to reduce friction during use. The radiopaque shaft and distal marker(s) facilitate fluoroscopic visualization. Device dimensions and configuration are shown on the product label. A rotating hemostatic Premarket Notification, Special 510(k) Trevo Trak 21 Microcatheter Stryker Neurovascular A rotating hemostatic valve with side-arm adapter is provided with each microcatheter. The Trevo Trak 21 Microcatheter provides vascular access to the neurovascular for selective placement of devices and/or the delivery of diagnostic agents into the neurovasculature during diagnostic and/or therapeutic endovascular procedures when used as intended. valve with side-arm adapter is provided with each microcatheter.
Accessory Devices Rotating Hemostasis Valve (RHV) packaged with device Same Outer Jacket Polymeric microcatheter Same Shaft Braid Stainless Steel Same Strain Relief Polyolefin Same Inner Layer PTFE Same Catheter Hub Polyurethane Same Marker Band Platinum/Iridium Same Adhesive Acrylic (Acrylated Urethane) Same Outer Jacket Coating Hydrophilic Coating Same Labeled Shaft Outer Diameter 2.4F/2.0F Same Labeled Shaft Inner Diameter 0.021” Same Effective Length 162 cm Same Packaging Materials HDPE Packaging Hoop, Tyvek/Film Pouch, SBS Carton Same Premarket Notification, Special 510(k) Trevo Trak 21 Microcatheter Stryker Neurovascular and Configuration Sterilization Method EO Sterilization Same How Supplied Single Use/Sterile Same Principles of Operation The device is advanced into the neurovascular vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the neurovasculature to the desired location. The device is advanced into the vasculature over an appropriately sized guide wire. Once the microcatheter is inserted, the catheter can be advanced through the vasculature to the desired location Shelf life 2-year Same Performance Testing The K261889 submission does not propose any modifications to the subject device design, materials, or manufacturing processes. The proposed changes to the intended use and indications for use, limiting device use to neurovasculature only, did not warrant new testing.
Shelf-Life Testing Shelf life testing previously conducted for the Trevo Trak 21 Microcatheter in K192122 was leveraged since there are no changes to device materials, design, or manufacturing processes as a result of the intended use, indications for use and labeling changes. As with the Predicate device, the Subject device is labeled with a 2-year shelf life.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。