HALYARD* ARMOR-FLEX* Sterilization Wrap with ASSURE-DRY* Technology; HALYARD* HEAVY-FLEX* Sterilization Wrap with ASSURE-DRY* Technology; HALYARD* STANDARD-FLEX* Sterilization Wrap with ASSURE-DRY* Technology; HALYARD* H300W Sterilization Wrap with ASSURE-DRY* Technology
O&M Halyard, Inc.
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
HALYARD* ARMOR-FLEX* Sterilization Wrap with ASSURE-DRY* Technology, HALYARD* HEAVY-FLEX* Sterilization Wrap with ASSURE-DRY* Technology, STANDARD-FLEX* Sterilization Wrap with ASSURE-DRY* Technology, and H300W Sterilization Wrap with ASSURE-DRY* Technology are intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:
• Pre-vacuum steam
• The wrap was validated for dry times of 20 minutes with the moisture managing amber layer on the inside, and for 30 minutes with the moisture managing amber layer on the outside.
• 270°F/132°C for 4 minutes
• 270°F/132°C for 10 minutes
• 273.2°F/134°C for 3 minutes
• 273.2°F/134°C for 18 minutes
• 275°F/135°C for 3 minutes
• STERIS V-PRO® Low Temperature Sterilization Systems. The wrap was validated to be effectively aerated during the pre-programmed cycles.
• STERIS V-PRO® 60 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS V-PRO® s2 (Lumen, Non-Lumen and Flexible Cycles)
• STERIS V-PRO® maX (Lumen, Non-Lumen and Flexible Cycles)
• STERIS V-PRO® maX 2 (Lumen, Non-Lumen and Flexible Cycles)
• Advanced Sterilization Products STERRAD® Sterilization System
• STERRAD® 100NX® (Standard Cycle)
• STERRAD® 100NX® with ALLClear® Technology (Standard Cycle)
Validated Loads1,2
Sterilization Cycle: Pre-Vacuum Cycle3 Wrapped Package Content Weight ARMOR-FLEX: 25 lbs Wrapped Package Content Weight HEAVY-FLEX: 17 lbs Wrapped Package Content Weight STANDARD-FLEX: 13 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: ASP STERRAD 100NX and 100NX with ALLClear Technology Standard Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 10.7 lbs Wrapped Package Content Weight HEAVY-FLEX: 10.7 lbs Wrapped Package Content Weight STANDARD-FLEX: 10.7 lbs Wrapped Package Content Weight H300W: 10.7 lbs
Sterilization Cycle: STERIS V-PRO maX, maX 2 Lumen Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 19.65 lbs Wrapped Package Content Weight HEAVY-FLEX: 17 lbs Wrapped Package Content Weight STANDARD-FLEX: 13 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: STERIS V-PRO 60, V-PRO s2 Lumen Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 11 lbs Wrapped Package Content Weight HEAVY-FLEX: 11 lbs Wrapped Package Content Weight STANDARD-FLEX: 11 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: STERIS V-PRO maX, maX 2 Non-Lumen Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 25 lbs Wrapped Package Content Weight HEAVY-FLEX: 17 lbs Wrapped Package Content Weight STANDARD-FLEX: 13 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: STERIS V-PRO 60, V-PRO s2 Non-Lumen Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 25 lbs Wrapped Package Content Weight HEAVY-FLEX: 17 lbs Wrapped Package Content Weight STANDARD-FLEX: 13 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: STERIS V-PRO maX, maX 2 Flexible Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 24 lbs Wrapped Package Content Weight HEAVY-FLEX: 17 lbs Wrapped Package Content Weight STANDARD-FLEX: 13 lbs Wrapped Package Content Weight H300W: 9 lbs
Sterilization Cycle: STERIS V-PRO 60, V-PRO s2 Flexible Cycle4 Wrapped Package Content Weight ARMOR-FLEX: 11 lbs Wrapped Package Content Weight HEAVY-FLEX: 11 lbs Wrapped Package Content Weight STANDARD-FLEX: 11 lbs Wrapped Package Content Weight H300W: 9 lbs
1 Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap.
2 It is recommended to not exceed the maximum wrapped package content weights.
3 Stainless steel instrument tray (~2 lbs.) containing lightweight medical instrumentation made of various metals and polymers, including hinged/mated and serrated/knurled surfaces or lumens (lumen and four (4) Huck towels, as shown in the information above. HALYARD Tray Liner and Transport Tray has been validated with the pre-vacuum steam cycle.
4 Reference Validation Load information is below.
Validation Load Composition ASP STERRAD® 100NX® and 100NX® with ALLClear Technology Standard Cycle Validation Load Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load: - An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.) Refer to the STERRAD® 100NX® Sterilizer User’s Guide and STERRAD® 100NX® with ALLClear® Technology Sterilizers User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e., 21.4 lbs. per load).
STERIS V-PRO maX and STERIS V-PRO maX2 Lumen Cycle Intended Load †Medical devices with the following configurations:
• Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
• Single, dual or triple channeled devices with stainless steel lumens with
• An inside diameter of 0.77 mm or larger and a length of 527 mm or shorter
• An inside diameter of 0.8 mm or larger and a length of 542 mm or shorter
• An inside diameter of 0.48 mm or larger and a length of 100 mm or shorter
• Dead end lumen with
• An inside diameter of 1.3 mm or larger and a length of 73 mm or shorter
• Rigid non-metallic lumen (such as those used in endoscope sheaths, take-apart forceps and trocars) with
• An inside diameter of 3 mm or larger and a length of 298 mm or shorter
• An inside diameter of 4 mm or larger and a length of 424 mm or shorter † Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two trays with silicone mats for a total weight of 19.65 lbs.
STERIS V-PRO 60 and s2 Lumen Cycle Intended Load Metal and non-metal medical devices including instruments with diffusion-restricted spaces (such as the hinged portion of forceps or scissors) and single, dual or triple channeled rigid/semi-rigid endoscopes, with the following configurations:
• Single or dual channeled devices with stainless steel lumens with:
• An inside diameter of 0.77 mm or larger and a length of 410 mm or shorter
• An inside diameter of 1.8 mm or larger and a length of 542 mm or shorter
• Triple channeled devices with stainless steel lumens with:
• An inside diameter of 1.2 mm or larger and a length of 275 mm or shorter
• An inside diameter of 1.8 mm or larger and a length of 310 mm or shorter
• An inside diameter of 2.8 mm or larger and a length of 317 mm or shorter Total load weight restricted to 11lbs.
STERIS V-PRO maX and STERIS V-PRO maX2 Non-Lumen Cycle Intended Load Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.
Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Note: The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs.
STERIS® V-PRO® 60 and s2 Non-Lumen Cycle Intended Load Metal and non-metal non-lumened medical devices including non-lumened rigid, semi-rigid and flexible endoscopes and medical devices with diffusion-restricted spaces such as the hinged portion of forceps or scissors. Total load weight is restricted to 25lbs.
STERIS V-PRO maX and STERIS V-PRO maX 2 Flexible Cycle Intended Load Non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.
Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations: Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load.* The flexible endoscopes may contain single or dual channel lumens that have:
• An inside diameter of 1 mm or larger and a length of 1050 mm or shorter OR
• One flexible endoscope with a light cord (if not integral to endoscope), endoscope accessories, mat and additional non-lumened instruments.** The flexible endoscope may contain single or dual channel lumens that have:
• An inside diameter of 1 mm or larger and a length of 1050 mm or shorter Additional instruments may include non-lumened or lumened medical devices with the following configurations: Single, dual or triple channel stainless steel lumen that have:
• An inside diameter of 0.48 mm or larger and a length of 100 mm or shorter * T he validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope). ** The validation studies were conducted with a flexible endoscope in a tray with endoscope accessories, silicone mat, light cord (if not integral to endoscope) and 5 stainless steel lumens. Also included in the load was a tra y w ith additional instruments and silicone mat for a total weight of 24 lbs. STER IS V-PRO 60 and s2 Flexible Cycle Intended Load Load 1: Single or dual channeled Flexible Surgical Endoscopes or Bronchoscopes with lumens that have:
• An inside diameter of 1 mm or larger and a length of 990 mm or shorter. Load 2*: Non-lumened instrument, instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors, and medical devices containing stainless steel lumens with the following dimensions:
• An inside diameter of 0.76 mm or larger and a length of 233 mm or shorter
• An inside diameter of 1.0 mm or larger and a length of 254 mm or shorter
• An inside diameter of 1.8 mm or larger and a length of 542 mm or shorter *Additional stainless steel lumens are restricted to 12 lumens Total load weight is restricted to 11 lbs.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。