FDA DEVICE FACTS
3M™ Attest™ Steam Chemical Integrator (1243A, 1243B, 1243RE, 1243RES)
Solventum US, LLC
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
T he 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
| Cycle Type | Temperature | Exposure Time |
|---|---|---|
| Gravity | 250°F/121°C | 30 minutes |
| Gravity | 270°F/132°C | 3 minutes |
| Gravity | 270°F/132°C | 4 minutes |
| Gravity | 270°F/132°C | 10 minutes |
| Gravity | 270°F/132°C | 15 minutes |
| Gravity | 270°F/132°C | 25 minutes |
| Gravity | 275°F/135°C | 3 minutes |
| Gravity | 275°F/135°C | 10 minutes |
| Dynamic Air Removal | 250°F/121°C | 30 minutes |
| Dynamic Air Removal | 270°F/132°C | 4 minutes |
| Dynamic Air Removal | 270°F/132°C | 5 minutes |
| Dynamic Air Removal | 270°F/132°C | 6 minutes |
M inimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer:
| 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C |
| 16.5 Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes |
| Dynamic Air Removal | 270°F/132°C | 7 minutes |
|---|---|---|
| Dynamic Air Removal | 270°F/132°C | 8 minutes |
| Dynamic Air Removal | 270°F/132°C | 9 minutes |
| Dynamic Air Removal | 270°F/132°C | 10 minutes |
| Dynamic Air Removal | 273°F/134°C | 3 minutes |
| Dynamic Air Removal | 273°F/134°C | 4 minutes |
| Dynamic Air Removal | 275°F/135°C | 3 minutes |
| Feature | Comparison | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Device Models | Identical | ||||||||
| Cycle Type | Temperature | Exposure Time | Cycle Type | Temperature | Exposure Time | ||||
| Gravity | 250°F/121°C | 30 minutes | Gravity | 250°F/121°C | 30 minutes | ||||
| Gravity | 270°F/132°C | 3 minutes | Gravity | 270°F/132°C | 3 minutes | ||||
| Gravity | 270°F/132°C | 4 minutes | Gravity | 270°F/132°C | 4 minutes | ||||
| Gravity | 270°F/132°C | 10 minutes | Gravity | 270°F/132°C | 10 minutes | ||||
| Gravity | 270°F/132°C | 15 minutes | Gravity | 270°F/132°C | 15 minutes | ||||
| Gravity | 270°F/132°C | 25 minutes | Gravity | 270°F/132°C | 25 minutes | ||||
| Gravity | 275°F/135°C | 3 minutes | Gravity | 275°F/135°C | 3 minutes | ||||
| Gravity | 275°F/135°C | 10 minutes | Gravity | 275°F/135°C | 10 minutes | ||||
| Dynamic Air Removal | 250°F/121°C | 30 minutes | Dynamic Air Removal | 250°F/121°C | 30 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 4 minutes | Dynamic Air Removal | 270°F/132°C | 4 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 5 minutes | Dynamic Air Removal | 270°F/132°C | 5 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 6 minutes | Dynamic Air Removal | 270°F/132°C | 6 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 7 minutes | Dynamic Air Removal | 270°F/132°C | 7 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 8 minutes | Dynamic Air Removal | 270°F/132°C | 8 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 9 minutes | Dynamic Air Removal | 270°F/132°C | 9 minutes | ||||
| Dynamic Air Removal | 270°F/132°C | 10 minutes | Dynamic Air Removal | 270°F/132°C | 10 minutes |
Tables marked from the FDA document were reconstructed automatically. Check the original FDA document (links at the end of this page) for authoritative values.
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。