Safety lancet (XL)
Tianjin Huahong Technology Co., Ltd.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The safety lancet is intended for capillary blood sampling.
| Item | Predicate device (K260191) | Proposed device | Comments |
|---|---|---|---|
| Product name | Safety Lancet (Model: XXXVII) | Safety Lancet (Model: XL) | Same |
| Product Code | FMK | FMK | Same |
| Regulation No. | 21 CFR § 878.4850 | 21 CFR § 878.4850 | Same |
| Class | II | II | Same |
| Prescription/over- the-counter use | Over-The-Counter Use | Over-The-Counter Use | Same |
| Indication for use | The safety lancet is intended for capillary blood sampling. | The safety lancet is intended for capillary blood sampling. | Same |
| Applicable user | Healthcare professional or lay person | Healthcare professional or lay person | Same |
| Reuse durability | Single use | Single use | Same |
| Sterilization method and SAL | Sterilized by Radiation SAL=10⁻⁶ | Sterilized by Radiation SAL=10⁻⁶ | Same |
| Manufacturing aspects | The injection molding of the needle core is completed in the purification workshop, the steel needle is wrapped in the needle core, sterilization of the needle core is outsourced, the subsequent process is the assembly and packaging of the safety lancet | The injection molding of the needle core is completed in the purification workshop, the steel needle is wrapped in the needle core, sterilization of the needle core is outsourced, the subsequent process is the assembly and packaging of the safety lancet | Same |
| Design and Functionality aspects | The needle tip is sterile, and the plastic part of the needle core serves as a sterile barrier for the needle tip. Set the penetration depth by turning the depth adjuster ring before use. When using, the cap needs to be twisted off first, after pressing the button, the safety needle is activated, and the needle tip penetrates the skin under the action of the spring, Subsequently, under the action of the retract spring, the needle tip retracts back into the safety needle. | The needle tip is sterile, and the plastic part of the needle core serves as a sterile barrier for the needle tip. Set the penetration depth by turning the depth adjuster ring before use. When using, the cap needs to be twisted off first, after pressing the button, the safety needle is activated, and the needle tip penetrates the skin under the action of the spring, Subsequently, under the action of the retract spring, the needle tip retracts back into the safety needle. | Same |
| Gauge range | 1.50±0.02mm(16G) 1.40±0.02mm(17G) 1.20±0.01mm(18G) 1.07±0.01mm(19G) 0.91±0.01mm(20G) 0.82±0.01mm(21G) 0.72±0.01mm(22G) 0.64±0.01mm(23G) 0.57±0.01mm(24G) 0.51±0.01mm(25G) 0.46±0.01mm(26G) 0.41±0.01mm(27G) 0.36±0.01mm(28G) 0.34±0.01mm(29G) 0.31±0.01mm(30G) 0.26±0.01mm(31G) 0.24±0.01mm(32G) 0.21±0.01mm(33G) 0.19±0.01mm(34G) 0.18±0.01mm(35G) 0.17±0.01mm(36G) 0.16±0.01mm(37G) 0.15±0.01mm(38G) | 1.50±0.02mm(16G) 1.40±0.02mm(17G) 1.20±0.01mm(18G) 1.07±0.01mm(19G) 0.91±0.01mm(20G) 0.82±0.01mm(21G) 0.72±0.01mm(22G) 0.64±0.01mm(23G) 0.57±0.01mm(24G) 0.51±0.01mm(25G) 0.46±0.01mm(26G) 0.41±0.01mm(27G) 0.36±0.01mm(28G) 0.34±0.01mm(29G) 0.31±0.01mm(30G) 0.26±0.01mm(31G) 0.24±0.01mm(32G) 0.21±0.01mm(33G) 0.19±0.01mm(34G) 0.18±0.01mm(35G) 0.17±0.01mm(36G) 0.16±0.01mm(37G) 0.15±0.01mm(38G) | Same |
| Shelf-life | 5 years | 5 years | |
| Materials of parts in contact with human body | Needle: stainless steel, silicone oil Housing, Button, Ring: ABS, PS, PP and Calcium powder | Needle: stainless steel, silicone oil Housing, Button, Ring: ABS, PS, PP and Calcium powder | Same |
| Biocompatibility | All materials are identical to those of the predicate device and have been evaluated in accordance with ISO 10993 series standards | All materials are identical to those of the predicate device and have been evaluated in accordance with ISO 10993 series standards | Same |
| Performance requirements | Performance requirements are unchanged, and verification testing confirms that the proposed device meets the same performance specifications as the predicate device | Performance requirements are unchanged, and verification testing confirms that the proposed device meets the same performance specifications as the predicate device | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |