RARA CRAFT DATABASE for FDA Medical Devices
510(k)K262526

FDA DEVICE FACTS

Safety lancet (XL)

Tianjin Huahong Technology Co., Ltd.

FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。

DECISION DATESeptember 13, 2026PRODUCT CODEFMK

Indications for use

The safety lancet is intended for capillary blood sampling.

Table 1 General Comparison of Safety Lancet
ItemPredicate device (K260191)Proposed deviceComments
Product nameSafety Lancet (Model: XXXVII)Safety Lancet (Model: XL)Same
Product CodeFMKFMKSame
Regulation No.21 CFR § 878.485021 CFR § 878.4850Same
ClassIIIISame
Prescription/over- the-counter useOver-The-Counter UseOver-The-Counter UseSame
Indication for useThe safety lancet is intended for capillary blood sampling.The safety lancet is intended for capillary blood sampling.Same
Applicable userHealthcare professional or lay personHealthcare professional or lay personSame
Reuse durabilitySingle useSingle useSame
Sterilization method and SALSterilized by Radiation SAL=10⁻⁶Sterilized by Radiation SAL=10⁻⁶Same
Manufacturing aspectsThe injection molding of the needle core is completed in the purification workshop, the steel needle is wrapped in the needle core, sterilization of the needle core is outsourced, the subsequent process is the assembly and packaging of the safety lancetThe injection molding of the needle core is completed in the purification workshop, the steel needle is wrapped in the needle core, sterilization of the needle core is outsourced, the subsequent process is the assembly and packaging of the safety lancetSame
Design and Functionality aspectsThe needle tip is sterile, and the plastic part of the needle core serves as a sterile barrier for the needle tip. Set the penetration depth by turning the depth adjuster ring before use. When using, the cap needs to be twisted off first, after pressing the button, the safety needle is activated, and the needle tip penetrates the skin under the action of the spring, Subsequently, under the action of the retract spring, the needle tip retracts back into the safety needle.The needle tip is sterile, and the plastic part of the needle core serves as a sterile barrier for the needle tip. Set the penetration depth by turning the depth adjuster ring before use. When using, the cap needs to be twisted off first, after pressing the button, the safety needle is activated, and the needle tip penetrates the skin under the action of the spring, Subsequently, under the action of the retract spring, the needle tip retracts back into the safety needle.Same
Gauge range1.50±0.02mm(16G) 1.40±0.02mm(17G) 1.20±0.01mm(18G) 1.07±0.01mm(19G) 0.91±0.01mm(20G) 0.82±0.01mm(21G) 0.72±0.01mm(22G) 0.64±0.01mm(23G) 0.57±0.01mm(24G) 0.51±0.01mm(25G) 0.46±0.01mm(26G) 0.41±0.01mm(27G) 0.36±0.01mm(28G) 0.34±0.01mm(29G) 0.31±0.01mm(30G) 0.26±0.01mm(31G) 0.24±0.01mm(32G) 0.21±0.01mm(33G) 0.19±0.01mm(34G) 0.18±0.01mm(35G) 0.17±0.01mm(36G) 0.16±0.01mm(37G) 0.15±0.01mm(38G)1.50±0.02mm(16G) 1.40±0.02mm(17G) 1.20±0.01mm(18G) 1.07±0.01mm(19G) 0.91±0.01mm(20G) 0.82±0.01mm(21G) 0.72±0.01mm(22G) 0.64±0.01mm(23G) 0.57±0.01mm(24G) 0.51±0.01mm(25G) 0.46±0.01mm(26G) 0.41±0.01mm(27G) 0.36±0.01mm(28G) 0.34±0.01mm(29G) 0.31±0.01mm(30G) 0.26±0.01mm(31G) 0.24±0.01mm(32G) 0.21±0.01mm(33G) 0.19±0.01mm(34G) 0.18±0.01mm(35G) 0.17±0.01mm(36G) 0.16±0.01mm(37G) 0.15±0.01mm(38G)Same
Shelf-life5 years5 years
Materials of parts in contact with human bodyNeedle: stainless steel, silicone oil Housing, Button, Ring: ABS, PS, PP and Calcium powderNeedle: stainless steel, silicone oil Housing, Button, Ring: ABS, PS, PP and Calcium powderSame
BiocompatibilityAll materials are identical to those of the predicate device and have been evaluated in accordance with ISO 10993 series standardsAll materials are identical to those of the predicate device and have been evaluated in accordance with ISO 10993 series standardsSame
Performance requirementsPerformance requirements are unchanged, and verification testing confirms that the proposed device meets the same performance specifications as the predicate devicePerformance requirements are unchanged, and verification testing confirms that the proposed device meets the same performance specifications as the predicate deviceSame
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same

Tables marked from the FDA document were reconstructed automatically. Check the original FDA document (links at the end of this page) for authoritative values.

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: October 7, 2026