Zenith® Iliac Branch
Cook, Inc.
FDA-sourced product information for medical-device development research. Use the official FDA sources at the end of this page when the original public document wording is needed.医療機器開発の調査用に、FDA の公開情報から取った製品情報を示しています。公開文書の原文が必要なときは、このページの最後にある FDA の公式な出典をご覧ください。
Indications for use
The Zenith® Iliac Branch, when used with the necessary additional components (Zenith AAA devices and a covered bridging stent), is indicated for endovascular treatment of aortoiliac or iliac aneurysms to preserve internal iliac arterial blood flow when the distal sealing site in the common iliac artery is insufficient for the AAA device alone and when the vessel morphology is suitable for repair, including:
• Common iliac artery diameter at the level of the internal iliac artery (luminal diameter) ≥16 mm
• Adequate iliac/femoral access compatible with a 20 French (7.7mm O.D.) introduction system
• Non-aneurysmal external iliac artery fixation segment distal to the aneurysm: o With a length of at least 20mm o With a diameter measured outer wall to outer wall of no greater than 11mm and no less than 8mm.
• Non-aneurysmal internal iliac artery segment distal to the aneurysm: o With a length of at least 10mm (with 20-30mm being preferred) o With a diameter measured outer wall to outer wall no greater than 10mm and no less than 7mm