Indications for use
The Relay®Pro Thoracic Stent -Graft System is indicated for the endovascular repair of fusiform aneurysms and saccular aneurysms/penetrating atherosclerotic ulcers in the descending thoracic aorta in patients having appropriate anatomy, including:
• Iliac or femoral access vessel morphology that is compatible with vascular access techniques, devices, and/or accessories;
• Non-aneurysmal aortic neck diameter in the range of 20 – 42 mm;
• Non-aneurysmal proximal aortic neck lengths of:
• 15 mm for the 24 – 28 mm device diameters (Bare Stent Configuration)
• 20 mm for the 30 – 38 mm device diameters (Bare Stent Configuration)
• 25 mm for the 40 – 46 mm device diameters (Bare Stent Configuration)
• 25 mm for the 24 – 38 mm device diameters ( Non-Bare Stent Configuration)
• 30 mm for the 40 – 46 mm device diameters ( Non-Bare Stent Configuration)
• Non-aneurysmal distal aortic neck lengths of:
• 25 mm for the 24 – 38 mm device diameters
• 30 mm for the 40 – 46 mm device diameters