FDA DEVICE FACTS
GORE TAG Thoracic Branch Endoprosthesis (TBE Device)
W. L. Gore & Associates, Inc.
Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。
Indications for use
The GORE® TAG® Thoracic Branch Endoprosthesis is indicated for endovascular repair of lesions of the descending thoracic aorta, while maintaining flow into the left subclavian artery, in patients who are at high risk for debranching subclavian procedures and have:
• Adequate iliac/femoral access
• Proximal Aortic Landing Zones: o For Isolated Lesion Patients: Proximal landing zone cannot be aneurysmal, dissected, heavily calcified, or heavily thrombosed o For Dissection Patients: Primary entry tear must be distal to
the left subclavian artery and the proximal extent of the landing zone must not be dissected o Aortic inner diameter range 16-42 mm o Proximal segment length (length from distal edge of left subclavian artery to mid left common carotid ostium) of at least 2.0-4.0 cm, depending on Aortic Component selection o Proximal covered length (measured from distal edge of left subclavian artery to distal edge of left common carotid artery ostium) of at least 15–36 mm, depending on Aortic Component selection o For patients with prior ascending aorta or aortic arch repair with a surgical graft: at least 2 cm landing zone proximal to the distal anastomosis
• Left Subclavian Landing Zone: o Landing zone cannot be aneurysmal, dissected, heavily thrombosed and severely tortuous (180 degree turn within the treated length) o Left subclavian artery inner diameter of 6–18 mm, depending on Side Branch Portal diameter selected o Left subclavian artery minimum length of 2.5–3.0 cm, depending on Side Branch Portal diameter selected
• Distal Landing Zone (Isolated Lesion Patients only) o Outer curve length must be ≥ 2 cm proximal to celiac artery o Aortic inner diameter range 16-42 mm o Non aneurysmal, dissected, heavily calcified, or heavily thrombosed landing zone o Native Aorta or previously placed GORE ® TAG® Conformable Thoracic Stent Graft
Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。
OFFICIAL FDA SOURCES
FDA Record and Public Document Status
Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。