RARA CRAFT DATABASE for FDA Medical Devices
PMAP210032S015

FDA DEVICE FACTS

GORE TAG Thoracic Branch Endoprosthesis (TBE Device)

W. L. Gore & Associates, Inc.

Verified FDA Record identity for medical-device development research. This overview uses only fields stated in the FDA Record and is not a Device Description.医療機器開発の調査用に、確認済みの FDA Record の識別情報を示しています。この概要は FDA Record に記載された項目だけによるもので、機器説明ではありません。

DECISION DATEApril 11, 2025PRODUCT CODESDZ

Indications for use

The GORE® TAG® Thoracic Branch Endoprosthesis is indicated for endovascular repair of lesions of the aortic arch and descending thoracic aorta, while maintaining flow into a single aortic arch branch vessel, in patients who have:

• Adequate iliac/femoral access

P210032/S015 FDA Summary of Safety and Effectiveness Data

• Proximal Aortic Landing Zones: o For Isolated Lesion Patients: Proximal landing zone cannot be aneurysmal, dissected, heavily calcified, or heavily thrombosed o For Dissection Patients: Primary entry tear must be distal to the target branch vessel and the proximal extent of the landing zone must not be dissected o Aortic inner diameter range 16-42 mm o Proximal segment length (length from distal edge of target branch vessel to the midpoint of any proximal branch vessel) of at least 2.0-4.0 cm, depending on Aortic Component selection o Proximal covered length (measured from distal edge of target branch vessel to the midpoint of any proximal branch vessel) of at least 15–36 mm, depending on Aortic Component selection o For patients with prior ascending aorta or aortic arch repair with a surgical graft: at least 2 cm landing zone proximal to the distal anastomosis

• Target Branch Vessel Landing Zone: o Landing zone cannot be aneurysmal, dissected, heavily thrombosed and severely tortuous (180 degree turn within the treated length) o Target branch vessel inner diameter of 6–18 mm, depending on Side Branch Portal diameter selected o Target branch vessel minimum length of 2.5–3.0 cm, depending on Side Branch Portal diameter selected

• Distal Landing Zone (Isolated Lesion Patients only) o Outer curve length must be ≥ 2 cm proximal to celiac artery o Aortic inner diameter range 16-42 mm o Cannot be aneurysmal, dissected, heavily calcified, or heavily thrombosed o Native Aorta or previously placed GORE ® TAG® Conformable Thoracic Stent Graft

See the full product information製品情報をすべて見るには

Device description, AI/software, PCCP, evaluation and limitations are available with an account: seven days of full access after free registration, then with the paid plan.機器の説明、AI・ソフトウェア、PCCP、評価、制限事項は、アカウントでご覧いただけます。無料登録から 7 日間はすべて利用でき、その後は有料プランでご利用いただけます。

Start free 7-day access7日間無料で利用するLog inログイン

OFFICIAL FDA SOURCES

FDA Record and Public Document Status

Record and document link states are stored separately.FDA Record と公開文書のリンクの状態は、別々に記録しています。

Information reviewed:情報の確認日: September 2, 2026