Happy Health's Ring-Based Blood Pressure Monitor Is Cleared Without Cuff Calibration
Several cuffless blood pressure monitors, most of them worn on the wrist, already hold 510(k) clearances, but most still begin with a cuff: the user calibrates once, and the device takes over from there. The Happy Health Blood Pressure Monitor (K254000, cleared September 3, 2026) has no calibration step. It derives systolic and diastolic pressure in mmHg from the pulse wave and other signals picked up at the finger, without demographic inputs and without memory of an earlier reading.
The device, as FDA's record defines it, is the software that computes blood pressure. The ring is one of the "compatible, FDA cleared" sensors that supply the signals, and for now the only one: the company's own Happy Ring.
FDA published draft guidance on clinical performance testing for cuffless blood pressure devices in January 2026, calling for at least 85 subjects, the ISO 81060-2 procedure and a mean error within ±5 mmHg. This submission reached FDA in August, after the draft.
RA Craft Database is an independent service and is not affiliated with or endorsed by the U.S. FDA. Articles are for general information on product development and regulatory research; they are not legal or regulatory advice for a specific device. Official FDA documents remain the authoritative source.